[AMMRL] ¹⁹F NMR Signal Detection

From: Muhammad Adil Raees via groups.io <adil.raees=lums.edu.pk_at_groups.io>
Date: Wed, 14 Oct 2026 04:38:17 -0700

Dear All,

A student has shared an FID (attached - PTh-BOD-DCP) from a previous measurement
of her sample at another NMR facility, showing a ¹⁹F signal at
approximately −130 ppm. She is now reacquiring the data using the same
sample with me; however, I have been unable to detect the corresponding signal
on our 600 MHz Bruker system. My FID is also attached for comparison.

I have tried increasing the number of scans and adjusting O1P, but the signal
remains undetectable.

I would appreciate your advice on possible causes and any further troubleshooting
steps you would recommend.

Best regards,
--
Muhammad Adil Raees, Ph.D.
Manager NMR Facility
The Central Laboratory
Syed Babar Ali School of Sciences and Engineering
Lahore University of Management Sciences
Opposite Sector-U, D.H.A. 54792, Lahore
+92-330-2866755
adil.raees_at_lums.edu.pk


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3854): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3854__;!!PvDODwlR4mBZyAb0!Q5A1z1x5xQ06l7N4rm_e3-GttTAKAD=
La2s0iZ13IyhEgj9Jn_HNlRR7dcQjAjgPlPehP25x3WAxlyHgHXN-Vu7kq4D-k$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121749994=
/7559972__;!!PvDODwlR4mBZyAb0!Q5A1z1x5xQ06l7N4rm_e3-GttTAKADLa2s0iZ13IyhEgj=
9Jn_HNlRR7dcQjAjgPlPehP25x3WAxlyHgHXN-VuwCaYXJJ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






Received on Wed Oct 14 2026 - 06:01:57 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST