Dear Muhammad,
The signal - which looks real - in the original spectrum is extremely weak,
19F is a very sensitive nucleus so this tells immediately it is a very
dilute analyte giving rise to the signal.
There is not much difference in the probes (BBOF vs. QNP) to account for the
lack of detection on your 600, and the field strength only gives something
like an 80% increase in sensitivity (depending on what formula you use).
So assuming the pulse width you use is more-or-less right and the probe is
sufficiently tuned and matched then the next possibility is saturation of
the signal.
The original spectrum had a total time for relaxation (acq + D1) of 2.4 sec
using a 30 degree pulse whereas you had a total time for relaxation of 1.5
sec using a 90 degree pulse. I would suggest allowing more time for relaxation
and/or reducing the pulse width and dispensing with the dummy scans as well
as increasing the number of scans even more.
Good luck,
K. Klika
________________________________________
> From: Muhammad Adil Raees via groups.io <adil.raees=lums.edu.pk_at_groups.io>
> Sent: 14 October 2026 14:38:17
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] ¹⁹F NMR Signal Detection
Dear All,
A student has shared an FID (attached - PTh-BOD-DCP) from a previous measurement
of her sample at another NMR facility, showing a ¹⁹F signal at
approximately −130 ppm. She is now reacquiring the data using the same
sample with me; however, I have been unable to detect the corresponding signal
on our 600 MHz Bruker system. My FID is also attached for comparison.
I have tried increasing the number of scans and adjusting O1P, but the signal
remains undetectable.
I would appreciate your advice on possible causes and any further troubleshooting
steps you would recommend.
Best regards,
--
Muhammad Adil Raees, Ph.D.
Manager NMR Facility
The Central Laboratory
Syed Babar Ali School of Sciences and Engineering
Lahore University of Management Sciences
Opposite Sector-U, D.H.A. 54792, Lahore
+92-330-2866755
adil.raees_at_lums.edu.pk
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3855): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3855__;!!PvDODwlR4mBZyAb0!WwopU6JCBexOaVZ28Us-OZcdC3JX9F=
SJQZwOQbH6Q22fel2t1vpSoeZ7VW2Mttu1yQouXr2gsKym0iUt7xzZdPaV$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121749994=
/7559972__;!!PvDODwlR4mBZyAb0!WwopU6JCBexOaVZ28Us-OZcdC3JX9FSJQZwOQbH6Q22fe=
l2t1vpSoeZ7VW2Mttu1yQouXr2gsKym0iUt78zFTPVF$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Oct 14 2026 - 06:51:51 MST