Hi Muhammad,

Assuming that reference samples give you an expected s/n and P90, I would try
a longer D1 (5 sec) and a lot more scans.

Also stick with zgig (as in spectrum 77), not zg (78).

Best Regards,

Jürgen

_______________________________________________

Jürgen Schulte, Ph. D.
NMR Specialist
Binghamton University
SN Building, RM 1024
25 Murray Hill Road
Vestal, NY 13850

Ph.: 607-777-4405 Fax: 607-777-4478
schulte_at_binghamton.edu
http://nmr.binghamton.edu


From: main_at_ammrl.groups.io . `On Behalf Of Muhammad Adil Raees via groups.io
> Sent: Wednesday, October 14, 2026 7:38 AM
> To: main_at_ammrl.groups.io
> Subject: [External Email] [AMMRL] ¹⁹F NMR Signal Detection

Dear All,

A student has shared an FID (attached - PTh-BOD-DCP) from a previous measurement
of her sample at another NMR facility, showing a ¹⁹F signal at
approximately −130 ppm. She is now reacquiring the data using the same
sample with me; however, I have been unable to detect the corresponding signal
on our 600 MHz Bruker system. My FID is also attached for comparison.

I have tried increasing the number of scans and adjusting O1P, but the signal
remains undetectable.

I would appreciate your advice on possible causes and any further troubleshooting
steps you would recommend.

Best regards,

--

Muhammad Adil Raees, Ph.D.
Manager NMR Facility
The Central Laboratory
Syed Babar Ali School of Sciences and Engineering
Lahore University of Management Sciences
Opposite Sector-U, D.H.A. 54792, Lahore
+92-330-2866755
adil.raees_at_lums.edu.pk


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3856): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3856__;!!PvDODwlR4mBZyAb0!S8Egt94F3JksvLFflAk4Sdm_QoMwKv=
aP5suhFuKtRhIniIu0ordjYqXiE6L0RqMgdPx6qSFj4bYyzh-5yvtZErU3vs9a$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121751290=
/7559972__;!!PvDODwlR4mBZyAb0!S8Egt94F3JksvLFflAk4Sdm_QoMwKvaP5suhFuKtRhIni=
Iu0ordjYqXiE6L0RqMgdPx6qSFj4bYyzh-5yvtZEkgRBMT7$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Oct 14 2026 - 07:06:04 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST