Date: Wed, 14 Oct 2026 14:04:32 +0000

P.S. In the original spectrum there is also a much stronger signal at -146 ppm t
hat is quite broad (easier to see it with, say, 10 Hz lb), and it could be that
there is a dynamic process going on giving rise to the two signals. Ostensibly
you are running at the same temperature so the higher field should help to observe
the signals in such a case, but maybe there is a change in sample precluding
this, e.g. precipitation of an aggregate which might be why the signal at -146 ppm
is broad or a faster exchange rate for whatever reason.

KK
________________________________________
> From: main_at_ammrl.groups.io on behalf of Karel Klika via groups.io
> Sent: 14 October 2026 16:40:11
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] ¹⁹F NMR Signal Detection

Dear Muhammad,

The signal - which looks real - in the original spectrum is extremely weak,
19F is a very sensitive nucleus so this tells immediately it is a very dilute
analyte giving rise to the signal.

There is not much difference in the probes (BBOF vs. QNP) to account for the
lack of detection on your 600, and the field strength only gives something
like an 80% increase in sensitivity (depending on what formula you use).
So assuming the pulse width you use is more-or-less right and the probe is
sufficiently tuned and matched then the next possibility is saturation of
the signal.
The original spectrum had a total time for relaxation (acq + D1) of 2.4 secusing
a 30 degree pulse whereas you had a total time for relaxation of 1.5 sec using
a 90 degree pulse. I would suggest allowing more time for relaxation and/or
reducing the pulse width and dispensing with the dummy scans as well as
increasing the number of scans even more.

Good luck,
K. Klika

________________________________________
> From: Muhammad Adil Raees via groups.io <adil.raees=lums.edu.pk_at_groups.io>
> Sent: 14 October 2026 14:38:17
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] ¹⁹F NMR Signal Detection

Dear All,

A student has shared an FID (attached - PTh-BOD-DCP) from a previous measurement
of her sample at another NMR facility, showing a ¹⁹F signal at
approximately −130 ppm. She is now reacquiring the data using the same
sample with me; however, I have been unable to detect the corresponding signal
on our 600 MHz Bruker system. My FID is also attached for comparison.

I have tried increasing the number of scans and adjusting O1P, but the signal
remains undetectable.

I would appreciate your advice on possible causes and any further troubleshooting
steps you would recommend.

Best regards,
--
Muhammad Adil Raees, Ph.D.
Manager NMR Facility
The Central Laboratory
Syed Babar Ali School of Sciences and Engineering
Lahore University of Management Sciences
Opposite Sector-U, D.H.A. 54792, Lahore
+92-330-2866755
adil.raees_at_lums.edu.pk








-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3857): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3857__;!!PvDODwlR4mBZyAb0!Roz0jVO4PFTvGqCQMBZYboppE3M1gu=
h-s3wsCPdwGifW6Xnzj37xnfmEbM3XO4v0ytKrTRBXGTXm6ZTZX2GIG9Gt$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121749994=
/7559972__;!!PvDODwlR4mBZyAb0!Roz0jVO4PFTvGqCQMBZYboppE3M1guh-s3wsCPdwGifW6=
Xnzj37xnfmEbM3XO4v0ytKrTRBXGTXm6ZTZX0t8xcf5$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Wed Oct 14 2026 - 07:06:16 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST