Hi Muhammad,
Regarding your troubleshooting for the missing 19F signal at approximately
-130 ppm, in addition to Karel's great suggestions on relaxation times and
pulse width, you might also consider incorporating a known concentration of
Trifluoroacetic acid (TFA, e.g., 0.001% v/v or ~130.6 µM) as an internal
reference standard.
Using TFA as an internal reference gives you a sharp singlet reference peak
(typically around -76.5 ppm) to immediately verify pulse calibration, receiver
gain settings, and overall 19F detectability in that sample matrix before
spending time on long scans for dilute analytes.
Best regards,
Vas
University of Pennsylvania
-=-=-=-=-=-=-=-=-=-=-=-