Hi Muhammed,

Slightly different angle to try.

It's difficulty not having the original proton to confirm equivalent sample
conditions. Since you didn't include it with the other attachments I presume
the student doesn't have it either. If they collected the data personally then
there's a lesson for them about always taking a proton. BUT IF someone else
from this other facility collected it then there's a good chance the proton
data exists, and is perhaps obtainable on request. Everyone, industry labs
included, collects a proton first thing. It's standard practice, good science,
and an important CYA tool for when things go wrong.

Kind regards,
Luke


Luke Fulton, PhD
CHEM BLDG R003
NMR Core Facility Director
Unit 3060
COR2E & Department of Chemistry
55 N Eagleville Road
University of Connecticut
Storrs, CT 06279


________________________________
> From: main_at_ammrl.groups.io on behalf of Muhammad Adil Raees via groups.io
> Sent: Wednesday, October 14, 2026 7:38 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] ¹⁹F NMR Signal Detection

Dear All,

A student has shared an FID (attached - PTh-BOD-DCP) from a previous measurement
of her sample at another NMR facility, showing a ¹⁹F signal at
approximately −130 ppm. She is now reacquiring the data using the same
sample with me; however, I have been unable to detect the corresponding signal
on our 600 MHz Bruker system. My FID is also attached for comparison.

I have tried increasing the number of scans and adjusting O1P, but the signal
remains undetectable.

I would appreciate your advice on possible causes and any further troubleshooting
steps you would recommend.

Best regards,
--
Muhammad Adil Raees, Ph.D.
Manager NMR Facility
The Central Laboratory
Syed Babar Ali School of Sciences and Engineering
Lahore University of Management Sciences
Opposite Sector-U, D.H.A. 54792, Lahore
+92-330-2866755
adil.raees_at_lums.edu.pk



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3859): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3859__;!!PvDODwlR4mBZyAb0!VkZIwKNiFN_HzjkLvq_-pHQ10oJzd2=
NkNSQv-t3o9yxoUyevH5K8hHQA2ByRnByy-cSH7SgNuKyew_wbuwlkODUWMRm5$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121749994=
/7559972__;!!PvDODwlR4mBZyAb0!VkZIwKNiFN_HzjkLvq_-pHQ10oJzd2NkNSQv-t3o9yxoU=
yevH5K8hHQA2ByRnByy-cSH7SgNuKyew_wbuwlkOHy8z8i-$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Oct 14 2026 - 08:22:28 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST