Hi Stuart,
We have seen similar errors previously on our SampleJets. While some of them
have come and gone with changing caps or re-calibrating the coordinates,
the most trouble we had was when the shuttle was getting trapped by the
sliders and resulting in sample loading / sample not found / etc. errors.
Eventually, it started decapitating sample tubes...
In the BST, there are four pistons that control the two sliders that hold
the shuttle in place. Inside these pistons there is a rubber o-ring that can
wear out and allow an internal leak that results in the pistons not firing
completely / slowly. The easiest way to check this is to disconnect the BST
and pressurize each of the four lines individually. The sliders should fire
freely when you apply air to just one of the inlets. If you see them moving
slowly or not at all, this would indicate that there is an issue with the
piston / o-ring.
We were able to find and buy replacement pistons for $20-30 (a few years ago).
Replacing the o-ring would also be a cheaper option. Bruker does not sell
these parts individually and would charge you the cost of a new BST...
SampleJet errors are not fun - I often find myself throwing everything at them
and then not knowing what was the actual problem. Good luck!
Thanks,
Dan
[cid:image001.png_at_01DD5BF8.D3501200]
Dan Conroy, PhD
Facility Manager, CBC NMR Facility
Research Scientist, CCIC Solid-State NMR
The Ohio State University
College of Arts & Sciences || Department of Chemistry & Biochemistry
083 CBEC Building, 151 W. Woodruff Ave., Columbus, OH 43210
Mobile: 708-612-2549 || conroy.120_at_osu.edu || u.osu.edu/cbcnmr/<mailto:conroy.120_at_osu.edu>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3860): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3860__;!!PvDODwlR4mBZyAb0!WGbbFnHV1jivnevdCEnSmkNm15ObZ4=
98-m5a3oBbGrNpoovtWmBlMMjNd-mQRKYCYgjfAyAo2tV1XcUHbiU_QP6CGK5a$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121731772=
/7559972__;!!PvDODwlR4mBZyAb0!WGbbFnHV1jivnevdCEnSmkNm15ObZ498-m5a3oBbGrNpo=
ovtWmBlMMjNd-mQRKYCYgjfAyAo2tV1XcUHbiU_QLWxSnf9$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Oct 14 2026 - 16:12:36 MST