Dear AMMRL,
I have the following ICON error while using the sample racks of our SampleJet:
SampleJet: Transport failed - Carrier stuck in BST, Sample Changer: Error
No sample found at position x
The follow ICON holder entry has the subsequent error:
Fatal error: 100 STOP_CHANGER BACS command failed
The error on the display of the SampleJet is:
Carrier stuck in BST
When I see this error, the shuttle is often stuck (at a partial tilt) at the
top of the bore of the instrument, with the sample already inside the shuttle.
There are also several samples earlier in batch that the SampleJet did not
attempt to run. The error in these cases is:
SampleJet: Transport failed - Air pressure low, Sample changer: Error: No
sample found at position x
If I power cycle the SampleJet and console, as well as going through steps
of the SampleJet recovery procedure and performing a coordinate/lift calibration,
then this does not fix the issue. I have swapped the gas supply to nitrogen
from compressed air, but the problem continues. I have also leak tested the
line, but there are no significant leaks. All other instruments on the same
compressed air line do not have issues ejecting/injecting samples. The
gripper and o-ring are also OK.
The strangest part still is that this problem went away for ~3 weeks, but
has now returned. The issue appears to be sporadic, perhaps 10 samples will
run before a failure, sometimes only one will run.
I have been told by a few other NMR FMs that it can be the caps not fitting
correctly. However, I recently tried a brand new bags of caps and got the
same error.
The system pressure is 4.9 bar (according to the service window, which is
accompanied by a green light).
The console is an AvIII-HD nanobay running TS.V.3.8.0 on Alma9.
Any advice or suggestions are greatly appreciated.
Kind Regards,
Stuart Elliott.
Dr. Stuart J. Elliott
NMR Facility Manager
Molecular Sciences Research Hub
Imperial College London
London W12 0BZ
United Kingdom
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3853): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3853__;!!PvDODwlR4mBZyAb0!T-_X0OUGyZTU38Pn84-sjkIgJy4-DK=
9ae06E99yQgAyB9Lcj91pQ77_GETQUQAsuT-xPNibWcqh30gZFrw3oir9S6UJb$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121731772=
/7559972__;!!PvDODwlR4mBZyAb0!T-_X0OUGyZTU38Pn84-sjkIgJy4-DK9ae06E99yQgAyB9=
Lcj91pQ77_GETQUQAsuT-xPNibWcqh30gZFrw3oig2mU0Tf$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Oct 13 2026 - 05:52:36 MST