Hi Emily
We have done this recently! I would go back to your local stockist for tmc
legs and quote them the serial number and they should be able provide spares.
One proviso when I did mine they managed to provide levers and brackets
that were magnetic so I had to retrieve the old levers and brackets to use
with the new valves. Also you may want to have a non magnetic knippex to
hand to undo some of the fixings. Your analysis of the problem is pretty well
right, the rubber seals inside go hard and stop working, unfortunately they
don’t sell on their own as it would be a much more ecologically
sound way to repair it, currently you have to replace the whole assembly
regards
Gary
Dr Gary S Thompson NMR Facility Manager
CCPN CoI & Working Group Member
Wellcome Trust Biomolecular NMR Facility
School of Natural Sciences
University of Kent, Canterbury, Kent, England, CT2 7NZ
☎:01227 82 7117
✉️: g.s.thompson_at_kent.ac.uk
orchid: orcid.org/0000-0001-9399-7636
On 12 Oct 2026, at 19:00, Emily Crull via groups.io wrote:
Hi everyone,
We recently noticed a dramatic hissing coming from one of the legs of our AS600.
I've attached a photo of the leg. The leak is coming from right under the brass
grooved piece. The amount of air and location makes me think that one of the
membranes has simply rotted out. Our model number is 60-2496255.
Does anyone have any recommendations on where to get a rebuild kit or the parts?
Tips on how best to do this would also be greatly appreciated.
Best,
Emily
<600MHz_leg.jpeg>
London and South East University Group, a charity and company limited by
guarantee registered in England (company number 00986729). Registered Office:
Old Royal Naval College, Park Row, Greenwich, London SE10 9LS
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3869): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3869__;!!PvDODwlR4mBZyAb0!RaH5MiVetS1z0Qf9qtCHdqAo4kXUVC=
ZOSRssaxUWJ6iLQt9n_8_Ht-e3bUG3fZeB9Z3pRBkPLV3AU4_cjzbWLv7h95NG$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121719964=
/7559972__;!!PvDODwlR4mBZyAb0!RaH5MiVetS1z0Qf9qtCHdqAo4kXUVCZOSRssaxUWJ6iLQ=
t9n_8_Ht-e3bUG3fZeB9Z3pRBkPLV3AU4_cjzbWLh_4MBGd$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Oct 15 2026 - 11:48:17 MST