From: Shore, Jay <jay.shore_at_sdstate.edu>
Date: Sun, 16 Jun 2024 20:19:53 +0000


Okay … I am at my wits’ ends … again … which was not a very for trip
for me … seems like I am always in the neighborhood.

I changed probes on our Avance 600 from a PATXI to PABBO so that we could do
a wider range of nuclei (e.g., Pt).

Unfortunately, the bbo probe has never been calibrated (power and pulse lengths).
So, I went about calibrating and thought that as long as I saved the new power
levels and pulse lengths in the prosol file that they would show up when I
went to iconnmr. That is what the manual says. I saved the parameters in the
prosol file and saved it for the probe (all solvents). Then, I started iconnmr
(routine spectroscopy), set up an experiment, looked at the parameters (for
pulse length and power) and they were not the same as in the prosol file. Does
anyone have any ideas of what I am missing? Does anyone have a good reference
for calibrating deuterium power and pulse lengths or changing probes.

I would appreciate any advice.
Thank you.
Jay







[South Dakota State University]
Jay Shore
Associate Professor
Chemistry & Biochemistry & Physics
Avera Health & Science 355, Box 2202
Brookings, SD 57007
P: (605) 688-6356
SDState.edu<https://www.sdstate.edu/ >





-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1333): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1333__;!!PvDODwlR4mBZyAb0!UPZSQ2HJR69ePhPNZR2O5WHPuvSB57=
5XjXQWePQojqJciym7dB7ohZXxZAitGUZH1s9uo08bXJfN8v_J-sNM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106713226=
/7559972__;!!PvDODwlR4mBZyAb0!UPZSQ2HJR69ePhPNZR2O5WHPuvSB575XjXQWePQojqJci=
ym7dB7ohZXxZAitGUZH1s9uo08bXJfN8h9bEy-5$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Sun Jun 16 2024 - 17:32:24 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST