Re: [AMMRL] edprosol conner

From: Todd Rappe via groups.io <rapp0006=umn.edu_at_groups.io>
Date: Sun, 16 Jun 2024 22:10:49 -0500

You might check if there are any solvent-specific calibration values saved
in your prosol. I believe if those are present, they supercede the "all
solvents" values.

On Sun, Jun 16, 2024, 7:32 PM Shore, Jay <jay.shore_at_sdstate.edu> wrote:

>
>
> Okay … I am at my wits’ ends … again … which was not a very for trip for
> me … seems like I am always in the neighborhood.
>
> I changed probes on our Avance 600 from a PATXI to PABBO so that we could
> do a wider range of nuclei (e.g., Pt).
>
> Unfortunately, the bbo probe has never been calibrated (power and pulse
> lengths). So, I went about calibrating and thought that as long as I saved
> the new power levels and pulse lengths in the prosol file that they would
> show up when I went to iconnmr. That is what the manual says. I saved the
> parameters in the prosol file and saved it for the probe (all solvents).
> Then, I started iconnmr (routine spectroscopy), set up an experiment,
> looked at the parameters (for pulse length and power) and they were not the
> same as in the prosol file. Does anyone have any ideas of what I am
> missing? Does anyone have a good reference for calibrating deuterium power
> and pulse lengths or changing probes.
>
> I would appreciate any advice.
>
> Thank you.
>
> Jay
>
>
> *Jay Shore*
> *Associate Professor*
> *Chemistry & Biochemistry & Physics*
>
> Avera Health & Science 355, Box 2202
> Brookings, SD 57007
>
> *P:* (605) 688-6356
> SDState.edu <https://www.sdstate.edu/>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1334): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1334__;!!PvDODwlR4mBZyAb0!VpT0wsBcYD2HE22fB3avZiSr6aTwpi=
y9QZSrsZ_Vaxza51iTfEIqIfrX-rk1yeugfX6cuT3vxYWVsTMT30VmFD7TBC8S$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106713226=
/7559972__;!!PvDODwlR4mBZyAb0!VpT0wsBcYD2HE22fB3avZiSr6aTwpiy9QZSrsZ_Vaxza5=
1iTfEIqIfrX-rk1yeugfX6cuT3vxYWVsTMT30VmFEtD-cW7$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Jun 17 2024 - 05:53:51 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST