Re: [AMMRL] Help with ideas for MRI liquid helium recovery

From: Karel Klika via groups.io <karkli=utu.fi_at_groups.io>
Date: Wed, 9 Sep 2026 15:02:09 +0000

Hi Mattias,

I would definitely vent the vacuum once the helium is gone otherwise the
pressure build up will be immense with a sizeable amount of condensed gas
(nitrogen or air).

For magnets, another way to speed up the helium boil off is to allow the liquid
nitrogen to boil off since once the liquid nitrogen is finished the boil off
rate will be something like 40-fold higher. The removal of the liquid
nitrogen can helped along by blowing in nitrogen gas (definitely nitrogen
this time to avoid condensation and accumulation of oxygen) into the nitrogen
port, but it does take a lot longer than you would think - hence in the
case I was involved in impatience got the better of people and we went for
the introduction of nitrogen gas into the vacuum, which went well but was
still not good enough for them so against my advice they switched over to
helium with the ensuing quench.

Best,
Karel
________________________________________
> From: main_at_ammrl.groups.io on behalf of Matthias Brandl via groups.io
> Sent: 09 September 2026 11:46:18
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Help with ideas for MRI liquid helium recovery

Hi Karel and Bert,

Wow, thanks, I will file "add helium to vacuum" under "bad ideas to
never try" then! I didn't realize that you only need so little to get
such a big effect.

Just out of curiosity and concerning the nitrogen method: After the
helium is gone, would you rather completely vent the vacuum right then
and there, or wait until the pressure increases naturally? In our case I
wanted to make sure the drop-off plate would actually drop off (some -
probably totally irrational - fear of it being too stuck or something;
for safety I also had a Technoflex valve on a T directly on the vacuum
port), but of course that runs the risk of blockages by water ice later
in the warm-up process...

Best,
Matthias



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3761): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3761__;!!PvDODwlR4mBZyAb0!T2Vo32QgU5w-cP-scoYJSeXeazzXJ_=
tAneC__wAJ9xdHY71r-ATYymM7Jdm4uRvYgPrlGgfpfok01nlNN_QGeQ1d$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121144451=
/7559972__;!!PvDODwlR4mBZyAb0!T2Vo32QgU5w-cP-scoYJSeXeazzXJ_tAneC__wAJ9xdHY=
71r-ATYymM7Jdm4uRvYgPrlGgfpfok01nlNN86S-ADl$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Wed Sep 09 2026 - 09:35:09 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST