Hi Karel and Bert,
Wow, thanks, I will file "add helium to vacuum" under "bad ideas to
never try" then! I didn't realize that you only need so little to get
such a big effect.
Just out of curiosity and concerning the nitrogen method: After the
helium is gone, would you rather completely vent the vacuum right then
and there, or wait until the pressure increases naturally? In our case I
wanted to make sure the drop-off plate would actually drop off (some -
probably totally irrational - fear of it being too stuck or something;
for safety I also had a Technoflex valve on a T directly on the vacuum
port), but of course that runs the risk of blockages by water ice later
in the warm-up process...
Best,
Matthias
On 08/09/2026 22:28, Spin-Doc wrote:
> Hi,
>
> as everyone is posting here (I already sent a DM to the requested
> address), some additions:
>
> As Karel correctly indicated, do NOT use helium gas to break the vacuum!
> Even traces of helium gas in the vacuum chamber will lead to all liquid
> helium to essentially flash-evaporate - it looks like a quench, even if
> the magnet is no longer on field…! No helium recovery system will be
> able to capture that. Like Karel I have personally witnessd it once
> (unplanned of course - I re-used a thin Tygon tube to let in nitrogen
> gas into the vacuum chamber - only it had been used for a helium refill
> some time before… SWOOSH!).
>
> With nitrogen gas it is somewhat better under control, but it's not
> linear either. Usually I let in a constant, tiny stream of nitrogen gas
> into the vacuum chamber via a needle valve. Put your ear to the vacuum
> can - it's usually enough when you *just* can har it…
> With this procedure you *must* make sure that the vacuum drop-off plate
> isn't obstructed let alone screwed in place - the magnet can explode
> otherwise because all the nitrogen frozen onto the helium can will turn
> into gas again as soon as the boiling point of nitrogen is reached...
>
> Clemens is right in that one can often transfer at least some (I've
> gotten something like 50% efficiency) of the liquid helium back into an
> empty but cold dewar - these should normally be available from the
> nearby helium recovery system. Best to talk to the recovery guys about
> scheduling this…
>
> Another thing - do I understand it correctly that these magnets are
> nitrogen-free and have a cryocooler sitting on top? That will complicate
> things a bit because the magnet pressure is usually maintained by the
> cryocontroller - so the magnet must be treated as if you were doing a
> helium refill (likely switch the compressor off during the operation -
> best to check the manual).
> In case more time (days?) is required between dewar fillings you'll need
> to switch the cooler back on again to not lose the helium.
>
> Cheers
> Bert
>
>> Am 08.09.2026 um 20:37 schrieb Karel Klika via groups.io
>> <karkli=utu.fi_at_groups.io>:
>>
>> Hi Amber,
>>
>> As Mattias suggested, you can feed in nitrogen gas (air is fine)
>> whereupon it will all freeze and the heat released will accelerate the
>> helium boil off. The process can be well controlled and I have
>> witnessed this (unfortunately not to the end, see below). Of course
>> the problem is the build up of nitrogen ice but you should be able to
>> calculate how much gas it is possible to solidify in the available space.
>>
>> However, do not use helium gas, the helium will obviously not freeze
>> and the thermal conduction at the higher pressure will very quickly
>> lead to a quench and the process cannot be controlled as the pressure
>> increase leading to a quench is ridiculously small. This unfortunately
>> is something I have witnessed.
>>
>> Kind regards,
>> K. Klika
>> ________________________________________
>> From:main_at_ammrl.groups.io
>> <mailto:main_at_ammrl.groups.io><main_at_ammrl.groups.io
>> <mailto:main_at_ammrl.groups.io>> on behalf of Matthias Brandl
>> viagroups.io <http://groups.io/><matthias.brandl=tum.de@groups.io
>> <mailto:matthias.brandl=tum.de_at_groups.io>>
>> Sent: 08 September 2026 20:05:15
>> To:main_at_ammrl.groups.io <mailto:main_at_ammrl.groups.io>
>> Subject: Re: [AMMRL] Help with ideas for MRI liquid helium recovery
>>
>> Hi Amber,
>>
>> I don't have any idea how to retrieve the liquid helium from the magnet
>> as a liquid, but I do have personal experience with two methods of
>> accelerating boil-off:
>>
>> Method 1:
>> We recently decommissioned an LTQ-MS instrument with a 7T magnet on our
>> campus. We introduced gaseous nitrogen into the insulation vacuum to
>> boil off the liquid helium into our recovery system at a controlled rate
>> (well, it was controlled until eventually the coil quenched, which was
>> expected in our case), and this worked quite well. We even stopped the
>> boil-off somewhere in the middle for a fire drill that had been
>> scheduled for the day.
>>
>> Since the overwhelming majority of the nitrogen should freeze to the
>> helium vessel, you are basically using its heat of deposition to boil
>> off the helium, so expect lots of nitrogen ice in the vacuum space of
>> the cryostat (that may be a problem). On the other hand, you can all but
>> stop the increased boiloff by interrupting the nitrogen supply.
>>
>> Method 2:
>> When Bruker warms up magnets for moving, they circulate gas from the
>> helium vessel through tubing/heat exchangers outside the cryostat. They
>> use what is basically a big diaphragm vacuum pump (I guess modified to
>> use all stages in parallel rather than in series for higher flow) to
>> withdraw helium from the top of the cryostat, warm it up through several
>> meters of corrugated vacuum hose, and put it back into the cryostat
>> through the syphon ordinarily used for cooling down the coil.
>>
>> This is again very controllable; switch off the pump to stop, reduce the
>> flow in the loop to slow down the boil-off. Though beware that with a
>> high flow, you run both the risk of freezing the pump you are using (if
>> the tubing used as a heat exchanger isn't long enough), and of
>> overwhelming the helium recovery system (at high pump flows, this method
>> can produce lots of gas FAST, and our recovery people were pretty mad at
>> us!).
>>
>>
>> Another alternative would be to contaminate the vacuum with helium, but
>> in this case you are adding heat transfer capacity which is not easily
>> removable should something happen to the recovery system in the
>> meantime. I don't have any experience with that method, nor have I heard
>> of anyone who tried it, but in principle it should work.
>>
>> Best,
>> Matthias
>>
>>
>> On 08/09/2026 15:23, Amber YaelSylvia Balazs via groups.io wrote:
>>> Hi Spinlanders,
>>> Our team in Gothenburg, Sweden, is decommissioning two Bruker B-C 47/40
>>> USR-TT MRI magnets. The magnets have been de-energized but still contain
>>> an estimated 1,600 L of liquid helium.
>>> The NMR team has a helium recovery system, but it is located remotely
>>> from the MRI facility. Running a temporary line to capture the boil-off
>>> would be impractical and, at the expected recovery rate, could take
>>> close to six months.
>>> We are exploring alternatives to avoid losing this helium, but the
>>> magnet design has made it difficult to identify practical recovery
>>> options. We would greatly appreciate any suggestions from others who
>>> have experience with MRI decommissioning or helium recovery in similar
>>> situations.
>>> If you have ideas or would be willing to discuss options, please contact
>>> Richard Lewis at Richard.J.Lewis_at_astrazeneca.com.
>>> Thank you,
>>> Amber
>>>
>>>
>>>
>>> ------------------------------------------------------------------------
>>>
>>> *Confidentiality Notice: *This message is private and may contain
>>> confidential and proprietary information. If you have received this
>>> message in error, please notify us and remove it from your system and
>>> note that you must not copy, distribute or take any action in reliance
>>> on it. Any unauthorized use or disclosure of the contents of this
>>> message is not permitted and may be unlawful.
>>>
>>>
>>
>> --
>> Dr. rer. nat. Matthias Brandl (he/him or they/them)
>>
>> Technical University of Munich
>> TUM School of Natural Sciences - BNMRZ
>> Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)
>>
>> Lichtenbergstr. 4 (postal)
>> Ernst-Otto-Fischer-Str. 2 (physical)
>> 85748 Garching b. Muenchen
>>
>> Phone +49 89 289-52617
>>
>> matthias.brandl_at_tum.de
>> https://www.sattlerlab.de/
>> https://www.bnmrz.org/
>>
>>
>>
>>
>>
>>
>>
>>
>
> image001.png***
>
> Magnetic resonance equipment, service, training & more
> ***
> Selling & buying used NMR equipment (all vendors)
> ***
> Dr. rer. nat. Bert Heise
> Owner, Applications Scientist
> Spin-Doc Bert Heise
> Kiefernweg 13
> 58239 Schwerte
> Germany
> Mobile: +49 171 5135315
> Email: bert.heise_at_spin-doc.net <mailto:bert.heise_at_spin-doc.net>
> ******
> Web:
> ***www.spin-doc.net <http://www.spin-doc.net>
>
> ***
> Representing:
> Stelar Fast Field Cycling NMR Relaxometers (www.stelar.it <http://
> www.stelar.it>)
> MestreLab NMR Software (www.mestrelab.com <http://www.mestrelab.com/>)
> Barthel HF Amplifiers (www.barthel-hf.de <http://www.barthel-hf.de>)
> ***
>
>
--
Dr. rer. nat. Matthias Brandl (he/him or they/them)
Technical University of Munich
TUM School of Natural Sciences - BNMRZ
Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)
Lichtenbergstr. 4 (postal)
Ernst-Otto-Fischer-Str. 2 (physical)
85748 Garching b. Muenchen
Phone +49 89 289-52617
matthias.brandl_at_tum.de
https://www.sattlerlab.de/
https://www.bnmrz.org/
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3759): https://ammrl.groups.io/g/main/message/3759
Mute This Topic: https://groups.io/mt/121144451/7559972
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://ammrl.groups.io/g/main/leave/12177410/7559972/1125599165/xyzzy [walt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Sep 09 2026 - 05:56:05 MST