Hi,
as everyone is posting here (I already sent a DM to the requested address)
some additions:
As Karel correctly indicated, do NOT use helium gas to break the vacuum! Even
traces of helium gas in the vacuum chamber will lead to all liquid helium
to essentially flash-evaporate - it looks like a quench, even if the magnet
is no longer on field…! No helium recovery system will be able to
capture that. Like Karel I have personally witnessd it once (unplanned of
course - I re-used a thin Tygon tube to let in nitrogen gas into the vacuum
chamber - only it had been used for a helium refill some time before… SWOOSH!).
With nitrogen gas it is somewhat better under control, but it's not linear
either. Usually I let in a constant, tiny stream of nitrogen gas into the
vacuum chamber via a needle valve. Put your ear to the vacuum can - it's
usually enough when you *just* can har it…
With this procedure you *must* make sure that the vacuum drop-off plate isn't
obstructed let alone screwed in place - the magnet can explode otherwise
because all the nitrogen frozen onto the helium can will turn into gas
again as soon as the boiling point of nitrogen is reached...
Clemens is right in that one can often transfer at least some (I've gotten
something like 50% efficiency) of the liquid helium back into an empty but
cold dewar - these should normally be available from the nearby helium recovery
system. Best to talk to the recovery guys about scheduling this…
Another thing - do I understand it correctly that these magnets are nitrogen-free
and have a cryocooler sitting on top? That will complicate things a bit because
the magnet pressure is usually maintained by the cryocontroller - so the magnet
must be treated as if you were doing a helium refill (likely switch the compressor
off during the operation - best to check the manual).
In case more time (days?) is required between dewar fillings you'll need to switch
the cooler back on again to not lose the helium.
Cheers
Bert
> Am 08.09.2026 um 20:37 schrieb Karel Klika via groups.io:
>
> Hi Amber,
>
> As Mattias suggested, you can feed in nitrogen gas (air is fine) whereupon
> it will all freeze and the heat released will accelerate the helium boil
> off. The process can be well controlled and I have witnessed this (unfortunately
> not to the end, see below). Of course the problem is the build up of nitrogen
> ice but you should be able to calculate how much gas it is possible to solidify
> in the available space.
>
> However, do not use helium gas, the helium will obviously not freeze and
> the thermal conduction at the higher pressure will very quickly lead to a
> quench and the process cannot be controlled as the pressure increase leading
> to a quench is ridiculously small. This unfortunately is something I have
> witnessed.
>
> Kind regards,
> K. Klika
> ________________________________________
> From: main_at_ammrl.groups.io on behalf of Matthias Brandl via groups.io
> Sent: 08 September 2026 20:05:15
> To: main_at_ammrl.groups.io <mailto:main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Help with ideas for MRI liquid helium recovery
>
> Hi Amber,
>
> I don't have any idea how to retrieve the liquid helium from the magnet
> as a liquid, but I do have personal experience with two methods of
> accelerating boil-off:
>
> Method 1:
> We recently decommissioned an LTQ-MS instrument with a 7T magnet on our
> campus. We introduced gaseous nitrogen into the insulation vacuum to
> boil off the liquid helium into our recovery system at a controlled rate
> (well, it was controlled until eventually the coil quenched, which was
> expected in our case), and this worked quite well. We even stopped the
> boil-off somewhere in the middle for a fire drill that had been
> scheduled for the day.
>
> Since the overwhelming majority of the nitrogen should freeze to the
> helium vessel, you are basically using its heat of deposition to boil
> off the helium, so expect lots of nitrogen ice in the vacuum space of
> the cryostat (that may be a problem). On the other hand, you can all but
> stop the increased boiloff by interrupting the nitrogen supply.
>
> Method 2:
> When Bruker warms up magnets for moving, they circulate gas from the
> helium vessel through tubing/heat exchangers outside the cryostat. They
> use what is basically a big diaphragm vacuum pump (I guess modified to
> use all stages in parallel rather than in series for higher flow) to
> withdraw helium from the top of the cryostat, warm it up through several
> meters of corrugated vacuum hose, and put it back into the cryostat
> through the syphon ordinarily used for cooling down the coil.
>
> This is again very controllable; switch off the pump to stop, reduce the
> flow in the loop to slow down the boil-off. Though beware that with a
> high flow, you run both the risk of freezing the pump you are using (if
> the tubing used as a heat exchanger isn't long enough), and of
> overwhelming the helium recovery system (at high pump flows, this method
> can produce lots of gas FAST, and our recovery people were pretty mad at
> us!).
>
>
> Another alternative would be to contaminate the vacuum with helium, but
> in this case you are adding heat transfer capacity which is not easily
> removable should something happen to the recovery system in the
> meantime. I don't have any experience with that method, nor have I heard
> of anyone who tried it, but in principle it should work.
>
> Best,
> Matthias
>
>
> On 08/09/2026 15:23, Amber YaelSylvia Balazs via groups.io wrote:
>> Hi Spinlanders,
>> Our team in Gothenburg, Sweden, is decommissioning two Bruker B-C 47/40
>> USR-TT MRI magnets. The magnets have been de-energized but still contain
>> an estimated 1,600 L of liquid helium.
>> The NMR team has a helium recovery system, but it is located remotely
>> from the MRI facility. Running a temporary line to capture the boil-off
>> would be impractical and, at the expected recovery rate, could take
>> close to six months.
>> We are exploring alternatives to avoid losing this helium, but the
>> magnet design has made it difficult to identify practical recovery
>> options. We would greatly appreciate any suggestions from others who
>> have experience with MRI decommissioning or helium recovery in similar
>> situations.
>> If you have ideas or would be willing to discuss options, please contact
>> Richard Lewis at Richard.J.Lewis_at_astrazeneca.com.
>> Thank you,
>> Amber
>>
>>
>>
>> ------------------------------------------------------------------------
>>
>
> --
> Dr. rer. nat. Matthias Brandl (he/him or they/them)
>
> Technical University of Munich
> TUM School of Natural Sciences - BNMRZ
> Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)
>
> Lichtenbergstr. 4 (postal)
> Ernst-Otto-Fischer-Str. 2 (physical)
> 85748 Garching b. Muenchen
>
> Phone +49 89 289-52617
>
> matthias.brandl_at_tum.de
> https://www.sattlerlab.de/
> https://www.bnmrz.org/
>
>
>
Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Representing:
Stelar Fast Field Cycling NMR Relaxometers (
http://www.stelar.it)
MestreLab NMR Software (
http://www.mestrelab.com)
Barthel HF Amplifiers (
http://www.barthel-hf.de)
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3758): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3758__;!!PvDODwlR4mBZyAb0!RoUELluH0x2LYk5DsjG0qhMddVDtBe=
7CEVglnO_IPpV2fqUG4GYxXfpBJ7069gNdUbDAnxohvmRHM-0Ybw6LAQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121144451=
/7559972__;!!PvDODwlR4mBZyAb0!RoUELluH0x2LYk5DsjG0qhMddVDtBe7CEVglnO_IPpV2f=
qUG4GYxXfpBJ7069gNdUbDAnxohvmRHM-076qhC3w$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Sep 08 2026 - 18:05:04 MST