Re: [AMMRL] Help with ideas for MRI liquid helium recovery

From: Karel Klika via groups.io <karkli=utu.fi_at_groups.io>
Date: Tue, 8 Sep 2026 18:37:52 +0000

Hi Amber,

As Mattias suggested, you can feed in nitrogen gas (air is fine) whereupon
it will all freeze and the heat released will accelerate the helium boil off.
The process can be well controlled and I have witnessed this (unfortunately
not to the end, see below). Of course the problem is the build up of nitrogen
ice but you should be able to calculate how much gas it is possible to solidify
in the available space.

However, do not use helium gas, the helium will obviously not freeze and the
thermal conduction at the higher pressure will very quickly lead to a quench
and the process cannot be controlled as the pressure increase leading to a
quench is ridiculously small. This unfortunately is something I have winessed.

Kind regards,
K. Klika
________________________________________
> From: main_at_ammrl.groups.io on behalf of Matthias Brandl via groups.io
> Sent: 08 September 2026 20:05:15
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Help with ideas for MRI liquid helium recovery

Hi Amber,

I don't have any idea how to retrieve the liquid helium from the magnet
as a liquid, but I do have personal experience with two methods of
accelerating boil-off:

Method 1:
We recently decommissioned an LTQ-MS instrument with a 7T magnet on our
campus. We introduced gaseous nitrogen into the insulation vacuum to
boil off the liquid helium into our recovery system at a controlled rate
(well, it was controlled until eventually the coil quenched, which was
expected in our case), and this worked quite well. We even stopped the
boil-off somewhere in the middle for a fire drill that had been
scheduled for the day.

Since the overwhelming majority of the nitrogen should freeze to the
helium vessel, you are basically using its heat of deposition to boil
off the helium, so expect lots of nitrogen ice in the vacuum space of
the cryostat (that may be a problem). On the other hand, you can all but
stop the increased boiloff by interrupting the nitrogen supply.

Method 2:
When Bruker warms up magnets for moving, they circulate gas from the
helium vessel through tubing/heat exchangers outside the cryostat. They
use what is basically a big diaphragm vacuum pump (I guess modified to
use all stages in parallel rather than in series for higher flow) to
withdraw helium from the top of the cryostat, warm it up through several
meters of corrugated vacuum hose, and put it back into the cryostat
through the syphon ordinarily used for cooling down the coil.

This is again very controllable; switch off the pump to stop, reduce the
flow in the loop to slow down the boil-off. Though beware that with a
high flow, you run both the risk of freezing the pump you are using (if
the tubing used as a heat exchanger isn't long enough), and of
overwhelming the helium recovery system (at high pump flows, this method
can produce lots of gas FAST, and our recovery people were pretty mad at
us!).


Another alternative would be to contaminate the vacuum with helium, but
in this case you are adding heat transfer capacity which is not easily
removable should something happen to the recovery system in the
meantime. I don't have any experience with that method, nor have I heard
of anyone who tried it, but in principle it should work.

Best,
Matthias


On 08/09/2026 15:23, Amber YaelSylvia Balazs via groups.io wrote:
> Hi Spinlanders,
> Our team in Gothenburg, Sweden, is decommissioning two Bruker B-C 47/40
> USR-TT MRI magnets. The magnets have been de-energized but still contain
> an estimated 1,600 L of liquid helium.
> The NMR team has a helium recovery system, but it is located remotely
> from the MRI facility. Running a temporary line to capture the boil-off
> would be impractical and, at the expected recovery rate, could take
> close to six months.
> We are exploring alternatives to avoid losing this helium, but the
> magnet design has made it difficult to identify practical recovery
> options. We would greatly appreciate any suggestions from others who
> have experience with MRI decommissioning or helium recovery in similar
> situations.
> If you have ideas or would be willing to discuss options, please contact
> Richard Lewis at Richard.J.Lewis_at_astrazeneca.com.
> Thank you,
> Amber
>
>
>
> ------------------------------------------------------------------------
>
> *Confidentiality Notice: *This message is private and may contain
> confidential and proprietary information. If you have received this
> message in error, please notify us and remove it from your system and
> note that you must not copy, distribute or take any action in reliance
> on it. Any unauthorized use or disclosure of the contents of this
> message is not permitted and may be unlawful.
>
>

--
Dr. rer. nat. Matthias Brandl (he/him or they/them)

Technical University of Munich
TUM School of Natural Sciences - BNMRZ
Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)

Lichtenbergstr. 4 (postal)
Ernst-Otto-Fischer-Str. 2 (physical)
85748 Garching b. Muenchen

Phone +49 89 289-52617

matthias.brandl_at_tum.de
https://www.sattlerlab.de/
https://www.bnmrz.org/








-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3757): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3757__;!!PvDODwlR4mBZyAb0!WUon4Exc9YhE7UO_bgdWNieNrDq6JA=
PrUjbdLqRdI6IXEt5hZSyTQ_6adIb3llfT1LwMn5IrB4NVYAw-QnKYAWUy$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121144451=
/7559972__;!!PvDODwlR4mBZyAb0!WUon4Exc9YhE7UO_bgdWNieNrDq6JAPrUjbdLqRdI6IXE=
t5hZSyTQ_6adIb3llfT1LwMn5IrB4NVYAw-QvZunpM2$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Tue Sep 08 2026 - 12:00:58 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST