HI Amber
on high resolution magnets it is possbile to transfer the liquid out to a
dewar. There will be some losses but a lot of the helium can be recovered.
I do not know if this is possible with the MRI magnet.
And the next question is what to do with the liquid heluim in the transport
dewar. If there is nobody nearby that has a need it can be difficult to reuse.
Regards
Clemens
On 9/8/2026 9:23 AM, Amber YaelSylvia Balazs via groups.io wrote:
Hi Spinlanders,
Our team in Gothenburg, Sweden, is decommissioning two Bruker B-C 47/40 USR-TT
MRI magnets. The magnets have been de-energized but still contain an estimated
1,600 L of liquid helium.
The NMR team has a helium recovery system, but it is located remotely from
the MRI facility. Running a temporary line to capture the boil-off would be
impractical and, at the expected recovery rate, could take close to six months.
We are exploring alternatives to avoid losing this helium, but the magnet design
has made it difficult to identify practical recovery options. We would greatly
appreciate any suggestions from others who have experience with MRI decommissioning
or helium recovery in similar situations.
If you have ideas or would be willing to discuss options, please contact Richard
Lewis at Richard.J.Lewis_at_astrazeneca.com<mailto:Richard.J.Lewis_at_astrazeneca.com>.
Thank you,
Amber
________________________________
Confidentiality Notice: This message is private and may contain confidential
and proprietary information. If you have received this message in error,
please notify us and remove it from your system and note that you must not
copy, distribute or take any action in reliance on it. Any unauthorized use
or disclosure of the contents of this message is not permitted and may be
unlawful.
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3756): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3756__;!!PvDODwlR4mBZyAb0!XKBY8zxLK4KefSLnTY_oAtjrDJSRue=
9zl_y1jTlSHEOgPIbwPqC7NNbcI0dydRc0C23FPx_NrHZCCXdrwY0893367zpo$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121144451=
/7559972__;!!PvDODwlR4mBZyAb0!XKBY8zxLK4KefSLnTY_oAtjrDJSRue9zl_y1jTlSHEOgP=
IbwPqC7NNbcI0dydRc0C23FPx_NrHZCCXdrwY0891gtTq5H$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!XKBY8zxLK4KefSLnT=
Received on Tue Sep 08 2026 - 10:45:49 MST