Re: [AMMRL] Help with ideas for MRI liquid helium recovery

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 9 Sep 2026 23:26:46 +0200

Hi,

another micro-snippet: I successfully insta-quenched several "Ol'Granddad"
magnets which were due for the scrapyard by filling a standard laboratory
glove with just a few mL of helium boiloff gas from the magnet itself (glove
still mostly flat). Hold the glove against the vacuum valve and briefly
open it. Insta-BANG! Quite shocking to see how little helium it needs…

The fastest way to get the nitrogen out is to insert a sufficiently thin tube
long enough to not vanish right away into the nitrogen can (…),
seal it with some lab film or the like, plug all other N2 ports except one
and adapt a nitrogen gas hose to the remaining port to slightly pressurize
the dewar to actually spray the liquid N2 out.

Fully supporting Mike's comment - that's precisely what I do as well.

;-)
Bert

> Am 09.09.2026 um 19:47 schrieb Richard Fitch via groups.io:
>
> Hi all,
>
> Just to pile on, I made the error of venting with helium when I decommissioned
> our old AC-250 system and it did exactly as described. I took down the magnet
> with diode, charging stick and the power supply on our AC console. I was
> certain the field was down as I stuck a wrench to the bottom of the magnet
> (not on or in the bore, of course). When it fell, I gave it another two hours
> to be sure. Then I dropped a paper clip down the bore and it fell to the floor,
> so I was quite sure the magnet was discharged.
>
> When I vented, I was a bit concerned about nitrogen freezing and risk of
> pressurization so I bled in from a helium cylinder (not very much either) and
> sure enough, the thing went off like twin volcanoes. By the time it died
> down, I had quite a cloud of condensation in the room that came down about
> 2 feet from the ceiling. Quite spectacular and surreal. It made me wonder o
> if I had hallucinated about the magnet being discharged. It was the only
> time I have seen a magnet frost all over on the outside. I left it until
> the next day when everything had warmed up and continued venting until the
> vacuum disk fell to the floor. I was still a bit nervous, but it was and
> uneventful and anticlimactic. I decided the old adage about God watching
> over fools, drunks and children was accurate.
>
> Best of luck,
> Rick
>
> Richard W. Fitch
> Professor, Department of Chemistry and Physics
> Councilor, Wabash Valley Section, American Chemical Society
> Indiana State University
> 600 Chestnut Street
> Science Building, Rm. S35E
> Terre Haute, IN 47809
> Phone: (812) 237-2244
> Fax: (812) 237-2232
>
> “Do an apparently simple experiment that gives you an important bit of information...”
> Julius Axelrod
>

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3763): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3763__;!!PvDODwlR4mBZyAb0!WEZtWHCKbTtAsFDHKcBe8EgYFOjhY2=
sPOtOAlhgfWkx-o04k2yWF-Ki5yLxpIXaOilbnDEfE39Wdxz28KuDDFw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121144451=
/7559972__;!!PvDODwlR4mBZyAb0!WEZtWHCKbTtAsFDHKcBe8EgYFOjhY2sPOtOAlhgfWkx-o=
04k2yWF-Ki5yLxpIXaOilbnDEfE39Wdxz1OKIZ4ug$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!WEZtWHCKbTtAsFDHK=
cBe8EgYFOjhY2sPOtOAlhgfWkx-o04k2yWF-Ki5yLxpIXaOilbnDEfE39Wdxz0B_7GXig$ [wa=
lt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Sep 09 2026 - 17:10:47 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST