Re: [AMMRL] Help with ideas for MRI liquid helium recovery

From: Jeff Walton via groups.io <jhwalton=ucdavis.edu_at_groups.io>
Date: Thu, 10 Sep 2026 23:05:48 +0000

Dear All,

One more caution. We had a drop-off plate fail on an Oxford 300 MHz wide bore.
I brought the magnet down and while still cold put it out in the hall to let
the cryogens boil off. We had other plans for the space. Several days later
when they did boil off the drop-off plate failed catastrophically at 4 in the
morning. Scared the you-know-what out of the janitor. We suspect that there
was some sort of small leak into the vacuum chamber and when it warmed up the
air-ice turned to vapor increasing the pressure. The drop-off plate did not
drop. Again I suspect the O-ring on the drop-off plate was corroded or deteriorated
or grease used acted more like glue. In any event, something prevented it from
dropping and when it finally did go, lots of noise and I am told the hall was
filled with mylar snow.

I’ve pulled a few magnets down and this was the only time it happened.
And I haven’t heard of anything like this happening to anyone else.
But it might be worthwhile to put a Tee on the vacuum port with a pop-off
valve and a shut off valve, and pump that down. Let the pop-off handle it
if the drop off plate doesn’t drop.

A cautionary tale. The only good thing about this episode is that no one got hurt.

Jeff

On Sep 9, 2026, at 2:26 PM, Spin-Doc via groups.io wrote:

Hi,

another micro-snippet: I successfully insta-quenched several "Ol'Granddad"
magnets which were due for the scrapyard by filling a standard laboratory
glove with just a few mL of helium boiloff gas from the magnet itself (glove
still mostly flat). Hold the glove against the vacuum valve and briefly open
it. Insta-BANG! Quite shocking to see how little helium it needs…

The fastest way to get the nitrogen out is to insert a sufficiently thin tube
long enough to not vanish right away into the nitrogen can (…), seal
it with some lab film or the like, plug all other N2 ports except one and
adapt a nitrogen gas hose to the remaining port to slightly pressurize the
dewar to actually spray the liquid N2 out.

Fully supporting Mike's comment - that's precisely what I do as well.

;-)
Bert

Am 09.09.2026 um 19:47 schrieb Richard Fitch via groups.io:

Hi all,

Just to pile on, I made the error of venting with helium when I decommissioned
our old AC-250 system and it did exactly as described. I took down the magnet
with diode, charging stick and the power supply on our AC console.

I was certain the field was down as I stuck a wrench to the bottom of the magnet
(not on or in the bore, of course). When it fell, I gave it another two hours
to be sure. Then I dropped a paper clip down the bore and it fell to the floor,
so I was quite sure the magnet was discharged.

When I vented, I was a bit concerned about nitrogen freezing and risk of
pressurization so I bled in from a helium cylinder (not very much either) and
sure enough, the thing went off like twin volcanoes. By the time it died down,
I had quite a cloud of condensation in the room that came down about 2 feet
from the ceiling. Quite spectacular and surreal. It made me wonder if I had
hallucinated about the magnet being discharged. It was the only time I have
seen a magnet frost all over on the outside. I left it until the next day when
everything had warmed up and continued venting until the vacuum disk fell to
the floor. I was still a bit nervous, but it was and uneventful and anticlimactic.
I decided the old adage about God watching over fools, drunks and children was accurate.

Best of luck,
Rick

Richard W. Fitch
Professor, Department of Chemistry and Physics
Councilor, Wabash Valley Section, American Chemical Society
Indiana State University
600 Chestnut Street
Science Building, Rm. S35E
Terre Haute, IN 47809
Phone: (812) 237-2244
Fax: (812) 237-2232

“Do an apparently simple experiment that gives you an important bit of information...”
Julius Axelrod

The love of money is the root of some evil. Fear and insecurity are the root of most of it.



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3766): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3766__;!!PvDODwlR4mBZyAb0!VzFPaxVXVf3TmsRgK4KZbM0xdExYYy=
rSVd2FK0BKJbpVasJx6XVtdInXpHJSeZtOV7Xbf72G1N68pS-2gez-pl3psYQN$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121144451=
/7559972__;!!PvDODwlR4mBZyAb0!VzFPaxVXVf3TmsRgK4KZbM0xdExYYyrSVd2FK0BKJbpVa=
sJx6XVtdInXpHJSeZtOV7Xbf72G1N68pS-2gez-pubBNsD0$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Sep 10 2026 - 16:21:25 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST