Re: [AMMRL] Help with ideas for MRI liquid helium recovery

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Thu, 10 Sep 2026 16:54:07 +0000 (UTC)

 Hi all,

Just to add to Bert's comment about getting the nitrogen out--we always sent
the output of that tube to some sort of storage vessel (dewar, large garbage
can, etc) so that you weren't spraying LN2 everywhere.  Usually we'd
use a long piece of latex tubing on the output to route the liquid to wherever.

And if you've got a piece of latex hose and a couple of zip ties, that works
well to seal the output hose to the dewar--put a length of latex hose around
your other tubing (we used to use copper tubing from home depot) and then when
you've fed the hose into the N2 dewar, you can put the latex hose around the
port like you're doing a fill, and use the zip ties to hold it around the
copper tubing.  If that makes sense.

Cheers,Mike

On Wednesday, September 9, 2026 at 06:08:33 PM MDT, Spin-Doc wrote:

 Hi,

another micro-snippet: I successfully insta-quenched several "Ol'Granddad"
magnets which were due for the scrapyard by filling a standard laboratory
glove with just a few mL of helium boiloff gas from the magnet itself (glove
still mostly flat). Hold the glove against the vacuum valve and briefly open
it. Insta-BANG! Quite shocking to see how little helium it needs… 

The fastest way to get the nitrogen out is to insert a sufficiently thin tube
long enough to not vanish right away into the nitrogen can (…),
seal it with some lab film or the like, plug all other N2 ports except one
and adapt a nitrogen gas hose to the remaining port to slightly pressurize
the dewar to actually spray the liquid N2 out.

Fully supporting Mike's comment - that's precisely what I do as well.

;-)Bert


Am 09.09.2026 um 19:47 schrieb Richard Fitch via groups.io:

Hi all,  Just to pile on, I made the error of venting with helium when '
I decommissioned our old AC-250 system and it did exactly as described. 
I took down the magnet with diode, charging stick and the power supply on
our AC console.  I was certain the field was down as I stuck a wrench
to the bottom of the magnet (not on or in the bore, of course).  When
it fell, I gave it another two hours to be sure.  Then I dropped a paper
clip down the bore and it fell to the floor, so I was quite sure the magnet was
discharged.
When I vented, I was a bit concerned about nitrogen freezing and risk of
pressurization so I bled in from a helium cylinder (not very much either) and
sure enough, the thing went off like twin volcanoes.  By the time it died down,
I had quite a cloud of condensation in the room that came down about 2 feet from
the ceiling.  Quite spectacular and surreal.  It made me wonder if I
had hallucinated about the magnet being discharged.  It was the only time
I have seen a magnet frost all over on the outside.  I left it until the
next day when everything had warmed up and continued venting until the vacuum
disk fell to the floor.  I was still a bit nervous, but it was and uneventful
and anticlimactic.  I decided the old adage about God watching over fools,
drunks and children was accurate.
Best of luck,
Rick

Richard W. Fitch
Professor, Department of Chemistry and Physics
Councilor, Wabash Valley Section,
American Chemical Society
Indiana State University
600 Chestnut Street
Science Building, Rm. S35E
Terre Haute, IN 47809
Phone: (812) 237-2244
Fax: (812) 237-2232

“Do an apparently simple experiment that gives you an important bit=
 of information...”
Julius Axelrod  
The love of money is the root of some evil.  Fear and insecurity are the root of most of it.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3765): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3765__;!!PvDODwlR4mBZyAb0!XwAt7rDGLVMb7u22ltufJb9UpPH-sZ=
mlPRPrctSVnxrESqHLvxAFjZoXaGpY0FvFm2UOjphnCZ3CIf2FIlUn5xD_7irt$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121144451=
/7559972__;!!PvDODwlR4mBZyAb0!XwAt7rDGLVMb7u22ltufJb9UpPH-sZmlPRPrctSVnxrES=
qHLvxAFjZoXaGpY0FvFm2UOjphnCZ3CIf2FIlUn5-Zzd7o9$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Sep 10 2026 - 11:10:09 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST