Hi all,
Just to pile on, I made the error of venting with helium when I decommissioned
our old AC-250 system and it did exactly as described. I took down the magnet
with diode, charging stick and the power supply on our AC console. I was
certain the field was down as I stuck a wrench to the bottom of the magnet (not
on or in the bore, of course). When it fell, I gave it another two hours to
be sure. Then I dropped a paper clip down the bore and it fell to the floor,
so I was quite sure the magnet was discharged.
When I vented, I was a bit concerned about nitrogen freezing and risk of
pressurization so I bled in from a helium cylinder (not very much either) and
sure enough, the thing went off like twin volcanoes. By the time it died
down, I had quite a cloud of condensation in the room that came down about 2
feet from the ceiling. Quite spectacular and surreal. It made me wonder if
I had hallucinated about the magnet being discharged. It was the only time
I have seen a magnet frost all over on the outside. I left it until the next
day when everything had warmed up and continued venting until the vacuum
disk fell to the floor. I was still a bit nervous, but it was and uneventful
and anticlimactic. I decided the old adage about God watching over fools,
drunks and children was accurate.
Best of luck,
Rick
Richard W. Fitch
Professor, Department of Chemistry and Physics
Councilor, Wabash Valley Section, American Chemical Society
Indiana State University
600 Chestnut Street
Science Building, Rm. S35E
Terre Haute, IN 47809
Phone: (812) 237-2244
Fax: (812) 237-2232
"Do an apparently simple experiment that gives you an important bit of information..."
Julius Axelrod
The love of money is the root of some evil. Fear and insecurity are the root of most of it.
> From: main_at_ammrl.groups.io On Behalf Of Michael Groves via groups.io
> Sent: Wednesday, September 9, 2026 9:52 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Help with ideas for MRI liquid helium recovery
Just my $0.02... After you get rid of the helium, since the innards of the
magnet are cold you want to prevent moisture or anything else that will freeze
from getting in. Normally if we're going to warm up a magnet we do fill the
vacuum space with N2, with a one-way valve on the vacuum port as you have it
set up. And we do fill it until the vacuum space reaches ambient pressure.
After that, I was trained to hold the drop off plate closed with something
rigid enough to keep it in place but not so strong that it will cause problems
if pressure does build up, (i.e. a piece of stiff packing foam or the equivalent)
because you don't want a lot of air getting into the magnet while it's drying.
And if the drop-off plate does open up while things are still warming up there's
a big hole for gunk to get in... But you do want to allow for the release of
pressure as things warm up and the gas expands.
Hopefully that makes sense--it did in my head...
Cheers,
Mike
On Wednesday, September 9, 2026 at 06:56:04 AM MDT, Matthias Brandl via groups.io wrote:
Hi Karel and Bert,
Wow, thanks, I will file "add helium to vacuum" under "bad ideas to
never try" then! I didn't realize that you only need so little to get
such a big effect.
Just out of curiosity and concerning the nitrogen method: After the
helium is gone, would you rather completely vent the vacuum right then
and there, or wait until the pressure increases naturally? In our case I
wanted to make sure the drop-off plate would actually drop off (some -
probably totally irrational - fear of it being too stuck or something;
for safety I also had a Technoflex valve on a T directly on the vacuum
port), but of course that runs the risk of blockages by water ice later
in the warm-up process...
Best,
Matthias
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3762): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3762__;!!PvDODwlR4mBZyAb0!Sl7FKy2Sj5vzAsMuCeEwz00xamif3-=
nkMHckFE1KMQfrcPOUA_HO2JUa4SYBVA-Es2sx3u-ncPKaXoYXD-fCiAsVJoST$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121144451=
/7559972__;!!PvDODwlR4mBZyAb0!Sl7FKy2Sj5vzAsMuCeEwz00xamif3-nkMHckFE1KMQfrc=
POUA_HO2JUa4SYBVA-Es2sx3u-ncPKaXoYXD-fCiDosBdzS$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Sep 09 2026 - 10:51:24 MST