Just my $0.02... After you get rid of the helium, since the innards
of the magnet are cold you want to prevent moisture or anything else that will
freeze from getting in. Normally if we're going to warm up a magnet we do fill
the vacuum space with N2, with a one-way valve on the vacuum port as you have
it set up. And we do fill it until the vacuum space reaches ambient
pressure. After that, I was trained to hold the drop off plate closed
with something rigid enough to keep it in place but not so strong that it
will cause problems if pressure does build up, (i.e. a piece of stiff packing
foam or the equivalent) because you don't want a lot of air getting into the
magnet while it's drying. And if the drop-off plate does open up while
things are still warming up there's a big hole for gunk to get in...
But you do want to allow for the release of pressure as things warm up and
the gas expands.
Hopefully that makes sense--it did in my head...
Cheers,Mike
On Wednesday, September 9, 2026 at 06:56:04 AM MDT, Matthias Brandl via groups.io
Hi Karel and Bert,
Wow, thanks, I will file "add helium to vacuum" under "bad ideas to
never try" then! I didn't realize that you only need so little to get
such a big effect.
Just out of curiosity and concerning the nitrogen method: After the
helium is gone, would you rather completely vent the vacuum right then
and there, or wait until the pressure increases naturally? In our case I
wanted to make sure the drop-off plate would actually drop off (some -
probably totally irrational - fear of it being too stuck or something;
for safety I also had a Technoflex valve on a T directly on the vacuum
port), but of course that runs the risk of blockages by water ice later
in the warm-up process...
Best,
Matthias
On 08/09/2026 22:28, Spin-Doc wrote:
> Hi,
> as everyone is posting here (I already sent a DM to the requested
> address), some additions:
> As Karel correctly indicated, do NOT use helium gas to break the vacuum!
> Even traces of helium gas in the vacuum chamber will lead to all liquid
> helium to essentially flash-evaporate - it looks like a quench, even if
> the magnet is no longer on field…! No helium recovery system will be
> able to capture that. Like Karel I have personally witnessd it once
> (unplanned of course - I re-used a thin Tygon tube to let in nitrogen
> gas into the vacuum chamber - only it had been used for a helium refill
> some time before… SWOOSH!).
> With nitrogen gas it is somewhat better under control, but it's not
> linear either. Usually I let in a constant, tiny stream of nitrogen gas
> into the vacuum chamber via a needle valve. Put your ear to the vacuum
> can - it's usually enough when you *just* can har it…
> With this procedure you *must* make sure that the vacuum drop-off plate
> isn't obstructed let alone screwed in place - the magnet can explode
> otherwise because all the nitrogen frozen onto the helium can will turn
> into gas again as soon as the boiling point of nitrogen is reached...
> Clemens is right in that one can often transfer at least some (I've
> gotten something like 50% efficiency) of the liquid helium back into an
> empty but cold dewar - these should normally be available from the
> nearby helium recovery system. Best to talk to the recovery guys about
> scheduling this…
> Another thing - do I understand it correctly that these magnets are
> nitrogen-free and have a cryocooler sitting on top? That will complicate
> things a bit because the magnet pressure is usually maintained by the
> cryocontroller - so the magnet must be treated as if you were doing a
> helium refill (likely switch the compressor off during the operation -
> best to check the manual).
> In case more time (days?) is required between dewar fillings you'll need
> to switch the cooler back on again to not lose the helium.
>
> Cheers
> Bert
>
>> Am 08.09.2026 um 20:37 schrieb Karel Klika via groups.io:
>>
>> Hi Amber,
>>
>> As Mattias suggested, you can feed in nitrogen gas (air is fine)
>> whereupon it will all freeze and the heat released will accelerate the
>> helium boil off. The process can be well controlled and I have
>> witnessed this (unfortunately not to the end, see below). Of course
>> the problem is the build up of nitrogen ice but you should be able to
>> calculate how much gas it is possible to solidify in the available space.
>>
>> However, do not use helium gas, the helium will obviously not freeze
>> and the thermal conduction at the higher pressure will very quickly
>> lead to a quench and the process cannot be controlled as the pressure
>> increase leading to a quench is ridiculously small. This unfortunately
>> is something I have witnessed.
>>
>> Kind regards,
>> K. Klika
>> ________________________________________
>>
> Magnetic resonance equipment, service, training & more
> ***
> Selling & buying used NMR equipment (all vendors)
> ***
> Dr. rer. nat. Bert Heise
> Owner, Applications Scientist
> Spin-Doc Bert Heise
> Kiefernweg 13
> 58239 Schwerte
> Germany
> Mobile: +49 171 5135315
> Email: bert.heise_at_spin-doc.net
> ******
>
> ***
> Representing:
> Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
> MestreLab NMR Software (http://www.mestrelab.com)
> Barthel HF Amplifiers (http://www.barthel-hf.de
> ***
>
>
--
Dr. rer. nat. Matthias Brandl (he/him or they/them)
Technical University of Munich
TUM School of Natural Sciences - BNMRZ
Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)
Lichtenbergstr. 4 (postal)
Ernst-Otto-Fischer-Str. 2 (physical)
85748 Garching b. Muenchen
Phone +49 89 289-52617
matthias.brandl_at_tum.de
https://www.sattlerlab.de/
https://www.bnmrz.org/
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3760): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3760__;!!PvDODwlR4mBZyAb0!WkfNWQvks8k5m4ltrEr_70Nq_q-eZe=
I28Hryx38wTEOXAF8xP2bvrg5WjaLg4oGXkJbs7FUMikog0ZuZqrf-HBnF1M9c$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121144451=
/7559972__;!!PvDODwlR4mBZyAb0!WkfNWQvks8k5m4ltrEr_70Nq_q-eZeI28Hryx38wTEOXA=
F8xP2bvrg5WjaLg4oGXkJbs7FUMikog0ZuZqrf-HJSP5pZc$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Sep 09 2026 - 09:42:32 MST