Just curious, what would you consider a "significant change" in helium
consumption?
We have a 700 that consistently boiled off ~2% of helium per week for
years. Over the past several months, that number has increased to over 3%
weekly.
_____________________________________
Todd Rappe
Minnesota NMR Center <nmr.umn.edu>
612-625-9147 (office)
651-321-TODD (rings office and mobile)
612-624-8892 (magnet lab)
On Fri, Aug 28, 2026 at 3:01 PM Anklin, Clemens via groups.io wrote:
> Rick
>
> if it ain't broke don't fix it. If a magnet shows no significant change
> in nitrogen and helium consumption it is good.
>
> In order to reseal you have to de-energize and warm up. Nobody will
> guarantee that it comes back up after this procedure. The chances are good
> but never 100%.
> The procedure also costs a lot of helium.
>
> With best regards
>
> Clemens
>
>
>
>
> On 8/28/2026 3:14 PM, Richard Fitch via groups.io wrote:
>
> Dear Spinlanders,
>
> I have a quandary on magnets and lifetimes. I have heard many times from
> the group on the excellent lifetime for magnets but am curious as to the
> experience of the group with refurbishing/resealing older magnets. We are
> batting around the idea and it would be useful to know everyone’s successes
> and failures with such things. I’m happy to post an anonymized summary if
> there is sufficient interest.
>
> All the best,
>
> Rick
>
>
>
> Richard W. Fitch
> Professor, Department of Chemistry and Physics
>
> Councilor, Wabash Valley Section, American Chemical Society
> Indiana State University
> 600 Chestnut Street
> Science Building, Rm. S35E
> Terre Haute, IN 47809
> Phone: (812) 237-2244
> Fax: (812) 237-2232
>
> “Do an apparently simple experiment that gives you an important bit of
> information...”
> Julius Axelrod
>
>
>
> The love of money is the root of some evil. Fear and insecurity are the
> root of most of it.
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3745): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3745__;!!PvDODwlR4mBZyAb0!XZrweEKsquDjYxdA5ogZVTvPtXjNP4=
PDfnVbsNKhBFdce16ktRFuUuvtWsCJhGCAVEiIyJ5E3-yX-cm81N8bxy4hPnyc$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120977275=
/7559972__;!!PvDODwlR4mBZyAb0!XZrweEKsquDjYxdA5ogZVTvPtXjNP4PDfnVbsNKhBFdce=
16ktRFuUuvtWsCJhGCAVEiIyJ5E3-yX-cm81N8bxz1VhabV$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Sep 03 2026 - 10:32:13 MST