Re: [AMMRL] Reseal on older magnets

From: Ulrike Werner-Zwanziger via groups.io <Ulli.Zwanziger=Dal.ca_at_groups.io>
Date: Fri, 4 Sep 2026 14:59:16 -0300

We just use long wood dowels from the hardware store for occasional N2 measurements.

Cheers,
Ulli

On Fri, Sep 4, 2026 at 1:38 PM Albers, Thomas via groups.io wrote:

> Goodwinds can supply fiberglass rods at 6 and 8 ft length:
> https://goodwinds.com/product-category/carbon-fiberglass/fiberglass/solid-fiberglass/
>
> The shipping surcharge for oversize packages may be painful, though.
>
> Thomas
>
> ------------------------------
> *From:* main_at_ammrl.groups.io on behalf of Juergen
> Schulte via groups.io <schulte=binghamton.edu_at_groups.io>
> *Sent:* Friday, September 4, 2026 09:46
> *To:* main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> *Subject:* Re: [AMMRL] Reseal on older magnets
>
>
> On our old small magnets, I used to use a garolite dip stick (McMaster
> *8526K11*) to measure the LN2 level. But I was never able to find one
> long enough for our big magnets.
>
> The Helium dip tube also works somewhat. When the tip touches the LN2
> surface, drops of nitrogen will shoot out of the open end.
>
> *Best Regards,*
>
> *Jürgen*
>
> *_______________________________________________*
>
> *Jürgen Schulte, Ph. D.
> *NMR Specialist
> *Binghamton University
> *SN Building, RM 1024*
> *25 Murray Hill Road
> *Vestal, NY 13850
>
> *Ph.: 607-777-4405 <(607)%20777-4405> Fax: 607-777-4478
> <(607)%20777-4478>*
>
> schulte_at_binghamton.edu
> http://chemiris.binghamton.edu/NMR/Schulte
>
>
> *From:* main_at_ammrl.groups.io *On Behalf Of *Michael Groves via groups.io
> *Sent:* Thursday, September 3, 2026 3:59 PM
> *To:* main_at_ammrl.groups.io
> *Subject:* [External Email] Re: [AMMRL] Reseal on older magnets
>
> Hi Todd,
>
> That can be hard to quantify without some way of measuring. One simple
> modification is to add a flow gauge in-line with the N2 exhaust. It
> doesn't even have to be calibrated to N2--as long as the flow is in the
> right range and you can check a number periodically you can see any
> developing trends.
>
> Cheers,
>
> Mike
>
> On Thursday, September 3, 2026 at 12:11:25 PM MDT, Todd Rappe via groups.io wrote:
>
> This instrument doesn't have an N2 level sensor, so it's tricky for me to
> gauge accurately, but it doesn't seem like we're consuming liquid nitrogen
> faster than I'm accustomed to.
>
> _____________________________________
> Todd Rappe
> Minnesota NMR Center <nmr.umn.edu>
> 612-625-9147 <(612)%20625-9147> (office)
> 651-321-TODD (rings office and mobile)
> 612-624-8892 <(612)%20624-8892> (magnet lab)
>
> On Thu, Sep 3, 2026, 1:03 PM Michael Groves via groups.io wrote:
>
> I would consider a 50% increase a significant change and would definitely
> keep an eye on that magnet to see if the trend continues. That does sound
> like the vacuum might be failing. Has the N2 boiloff increased as well?
>
> Cheers,
>
> Mike
>
> On Thursday, September 3, 2026 at 11:32:12 AM MDT, Todd Rappe via groups.io wrote:
>
> Just curious, what would you consider a "significant change" in helium
> consumption?
>
> We have a 700 that consistently boiled off ~2% of helium per week for
> years. Over the past several months, that number has increased to over 3%
> weekly.
>
> _____________________________________
> Todd Rappe
> Minnesota NMR Center <nmr.umn.edu>
> 612-625-9147 <(612)%20625-9147> (office)
> 651-321-TODD (rings office and mobile)
> 612-624-8892 <(612)%20624-8892> (magnet lab)
>
--
Dr. Ulrike Werner-Zwanziger
Adjunct Professor
Department of Chemistry, NMR-3,
Dalhousie University
6245 Castine Way
PO BOX 15000
Halifax, NS B3H 4R2

e-mail: Ulli.Zwanziger_at_Dal.ca
Web: http://www.dal.ca/diff/nmr3.html
Tel. 1 902 494 8085
FAX: 1 902 494 1310


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3753): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3753__;!!PvDODwlR4mBZyAb0!RFfRLTsbNEa9xdT8LyjvzJoD6YCx7M=
M6TIf8TP5nyHllPmtlVvMDUv3SXmKrJS-sjjP0tUKxa5YBSuAtI9GJWuP3R4ZI$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120977275=
/7559972__;!!PvDODwlR4mBZyAb0!RFfRLTsbNEa9xdT8LyjvzJoD6YCx7MM6TIf8TP5nyHllP=
mtlVvMDUv3SXmKrJS-sjjP0tUKxa5YBSuAtI9GJWvBI3UT6$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Sep 04 2026 - 11:33:14 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST