Re: [AMMRL] Reseal on older magnets

From: Albers, Thomas via groups.io <talbers=nmhu.edu_at_groups.io>
Date: Fri, 4 Sep 2026 16:33:06 +0000

Goodwinds can supply fiberglass rods at 6 and 8 ft length:
https://goodwinds.com/product-category/carbon-fiberglass/fiberglass/solid-fiberglass/

The shipping surcharge for oversize packages may be painful, though.

Thomas

________________________________
> From: main_at_ammrl.groups.io on behalf of Juergen Schulte via groups.io
> Sent: Friday, September 4, 2026 09:46
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Reseal on older magnets

On our old small magnets, I used to use a garolite dip stick (McMaster 8526K11)
to measure the LN2 level. But I was never able to find one long enough for our
big magnets.

The Helium dip tube also works somewhat. When the tip touches the LN2 surface,
drops of nitrogen will shoot out of the open end.

Best Regards,

Jürgen

_______________________________________________

Jürgen Schulte, Ph. D.
NMR Specialist
Binghamton University
SN Building, RM 1024
25 Murray Hill Road
Vestal, NY 13850

Ph.: 607-777-4405 Fax: 607-777-4478

schulte_at_binghamton.edu

http://chemiris.binghamton.edu/NMR/Schulte



> From: main_at_ammrl.groups.io On Behalf Of Michael Groves via groups.io
> Sent: Thursday, September 3, 2026 3:59 PM
> To: main_at_ammrl.groups.io
> Subject: [External Email] Re: [AMMRL] Reseal on older magnets

Hi Todd,

That can be hard to quantify without some way of measuring. One simple
modification is to add a flow gauge in-line with the N2 exhaust. It doesn't
even have to be calibrated to N2--as long as the flow is in the right range
and you can check a number periodically you can see any developing trends.

Cheers,

Mike

On Thursday, September 3, 2026 at 12:11:25 PM MDT, Todd Rappe via groups.io wrote:

This instrument doesn't have an N2 level sensor, so it's tricky for me to
gauge accurately, but it doesn't seem like we're consuming liquid nitrogen
faster than I'm accustomed to.

_____________________________________
Todd Rappe
Minnesota NMR Center
nmr.umn.edu
612-625-9147 (office)
651-321-TODD (rings office and mobile)
612-624-8892 (magnet lab)

On Thu, Sep 3, 2026, 1:03 PM Michael Groves via groups.io wrote:

I would consider a 50% increase a significant change and would definitely keep
an eye on that magnet to see if the trend continues. That does sound like
the vacuum might be failing. Has the N2 boiloff increased as well?

Cheers,

Mike

On Thursday, September 3, 2026 at 11:32:12 AM MDT, Todd Rappe via groups.io wrote:

Just curious, what would you consider a "significant change" in helium consumption?

We have a 700 that consistently boiled off ~2% of helium per week for years.
Over the past several months, that number has increased to over 3% weekly.

_____________________________________
Todd Rappe
Minnesota NMR Center
nmr.umn.edu
612-625-9147 (office)
651-321-TODD (rings office and mobile)
612-624-8892 (magnet lab)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3752): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3752__;!!PvDODwlR4mBZyAb0!U6OUf3RpGtvUPqJX6QDMhkZk_jtgHa=
KRdkCyTK5P6lflf3Q94zBxh1OFWBWup11ZeZeBLhdiVrTBh4Ue6ZXLUUvoQ0IU$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120977275=
/7559972__;!!PvDODwlR4mBZyAb0!U6OUf3RpGtvUPqJX6QDMhkZk_jtgHaKRdkCyTK5P6lflf=
3Q94zBxh1OFWBWup11ZeZeBLhdiVrTBh4Ue6ZXLUUuhdbId$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Sep 04 2026 - 09:38:09 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST