Hi Todd
I second Mike Groves' comments. But it does not necessarily need to be the
seals. It can also be that the getter (small device with charcoal to absorb
gases) is saturated.
That would also lead to an increase. In either case the system would have to
be warmed up to either change the seals or the getter or usually both.
Clemens
On 9/3/2026 11:45 AM, Todd Rappe via groups.io wrote:
Just curious, what would you consider a "significant change" in helium consumption?
We have a 700 that consistently boiled off ~2% of helium per week for years.
Over the past several months, that number has increased to over 3% weekly.
_____________________________________
Todd Rappe
Minnesota NMR Center <nmr.umn.edu>
612-625-9147 (office)
651-321-TODD (rings office and mobile)
612-624-8892 (magnet lab)
On Fri, Aug 28, 2026 at 3:01 PM Anklin, Clemens via groups.io wrote:
Rick
if it ain't broke don't fix it. If a magnet shows no significant change in
nitrogen and helium consumption it is good.
In order to reseal you have to de-energize and warm up. Nobody will guarantee
that it comes back up after this procedure. The chances are good but never 100%.
The procedure also costs a lot of helium.
With best regards
Clemens
On 8/28/2026 3:14 PM, Richard Fitch via groups.io wrote:
Dear Spinlanders,
I have a quandary on magnets and lifetimes. I have heard many times from the
group on the excellent lifetime for magnets but am curious as to the experience
of the group with refurbishing/resealing older magnets. We are batting around
the idea and it would be useful to know everyone’s successes and
failures with such things. I’m happy to post an anonymized summary if
there is sufficient interest.
All the best,
Rick
Richard W. Fitch
Professor, Department of Chemistry and Physics
Councilor, Wabash Valley Section, American Chemical Society
Indiana State University
600 Chestnut Street
Science Building, Rm. S35E
Terre Haute, IN 47809
Phone: (812) 237-2244<tel:(812)%20237-2244>
Fax: (812) 237-2232<tel:(812)%20237-2232>
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3748): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3748__;!!PvDODwlR4mBZyAb0!XxJIRcPu2gOfXz8AIo0sj0guGcpIMb=
ayZP4fR69dAVkl5kv_dJdyDlE98C1x2ehvyqZ0Eyi6qSDvtORN1GYPO9ECBlCq$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120977275=
/7559972__;!!PvDODwlR4mBZyAb0!XxJIRcPu2gOfXz8AIo0sj0guGcpIMbayZP4fR69dAVkl5=
kv_dJdyDlE98C1x2ehvyqZ0Eyi6qSDvtORN1GYPO1eFotgj$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Sep 03 2026 - 11:41:03 MST