I would consider a 50% increase a significant change and would definitely
keep an eye on that magnet to see if the trend continues.  That does
sound like the vacuum might be failing.  Has the N2 boiloff increased
as well?

Cheers,Mike

On Thursday, September 3, 2026 at 11:32:12 AM MDT, Todd Rappe via groups.io wrote:

Just curious, what would you consider a "significant change" in helium consumption? 
We have a 700 that consistently boiled off ~2% of helium per week for years. Over the
past several months, that number has increased to over 3% weekly.

Todd Rappe
Minnesota NMR Center <nmr.umn.edu>
612-625-9147 (office)
651-321-TODD (rings office and mobile)
612-624-8892 (magnet lab)




On Fri, Aug 28, 2026 at 3:01 PM Anklin, Clemens via groups.io wrote:

Rick

if it ain't broke don't fix it.  If a magnet shows no significant change
in nitrogen and helium consumption it is good. 

In order to reseal you have to de-energize and warm up. Nobody will guarantee
that it comes back up after this procedure. The chances are good but never 100%.
The procedure also costs a lot of helium.

With best regards

Clemens


On 8/28/2026 3:14 PM, Richard Fitch via groups.io wrote:

Dear Spinlanders,

I have a quandary on magnets and lifetimes.  I have heard many times from
the group on the excellent lifetime for magnets but am curious as to the
experience of the group with refurbishing/resealing older magnets.  We
are batting around the idea and it would be useful to know everyone’s
successes and failures with such things.  I’m happy to post an
anonymized summary if there is sufficient interest.


All the best,

Rick


Richard W. Fitch
Professor, Department of Chemistry and Physics

Councilor, Wabash Valley Section, American Chemical Society
Indiana State University
600 Chestnut Street
Science Building, Rm. S35E
Terre Haute, IN 47809
Phone: (812) 237-2244
Fax: (812) 237-2232


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3746): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3746__;!!PvDODwlR4mBZyAb0!XMSyBP5j5v1xxQ_ywfac0ARbQg18AA=
Y4YfMqzPJLeqLeolhf6mDNN9pfrFZshmli51wfYDJ_WCKyWNBdMQyxP5vyyiNz$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120977275=
/7559972__;!!PvDODwlR4mBZyAb0!XMSyBP5j5v1xxQ_ywfac0ARbQg18AAY4YfMqzPJLeqLeo=
lhf6mDNN9pfrFZshmli51wfYDJ_WCKyWNBdMQyxP5rymLGD$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Sep 03 2026 - 11:03:24 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST