Re: [AMMRL] #JEOL JEOL ECS-400 autotune issue

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Thu, 3 Sep 2026 01:21:50 +0200

Hi Piotr,

did you also try out the probe of that other ECS400 system whether it shows
the same issue?

When you say you can tune manually to 31P that means you can set the coarse
tune knob to "X" by hand but it doesn't rotate there during auto-tuning?
Does the coarse motor rotate that axle at all during (pre)tuning?
Do both probes feel similar in terms of mechanical resistance/torque when turning the coarse knob by hand?

Cheers
Bert

> Am 02.09.2026 um 16:21 schrieb Piotr Dobrowolski via groups.io:
>
> I forgot to add that we tried to use another autotune controller module
> (including the shafts) from an ECS-400 console we decommissioned a few months
> ago. We've never had such a problem with that instrument. However, changing
> the controller did not fix the issue.
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3744): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3744__;!!PvDODwlR4mBZyAb0!UBl4nnuwOqY4yYKjhtwsQmtWbpwwxE=
McdRmqwo9mFbYYwgRBkWsfIL10xO2MQThKdWXMFSbwr2UZaA7onMZLsw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121043236=
/7559972__;!!PvDODwlR4mBZyAb0!UBl4nnuwOqY4yYKjhtwsQmtWbpwwxEMcdRmqwo9mFbYYw=
gRBkWsfIL10xO2MQThKdWXMFSbwr2UZaA7yFhocKw$
Mute #jeol:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/muteh=
ashtag/jeol__;!!PvDODwlR4mBZyAb0!UBl4nnuwOqY4yYKjhtwsQmtWbpwwxEMcdRmqwo9mFb=
YYwgRBkWsfIL10xO2MQThKdWXMFSbwr2UZaA6kx4uk1Q$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Sep 02 2026 - 17:11:35 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST