Dear Weilin,
what do they (not) do…?
Don't switch on (check power supply, especially capacitors)?
Not cooling (possibly refrigerant - difficult to refill the older units which
I believe still use chloro-/fluorocarbons)?
No airflow (ice block - check dew point of VT gas)?
Not counter-heating (broken cable within black air tubing - give it a wiggle
and twist, does it resume heating? Difficult to repair as the metal end tube
is full of expanded PU foam…)?
Unfortunately I don't have an FTS in PDF form - should scan it once...
Cheers
Bert
> Am 02.09.2026 um 21:50 schrieb Weilin Sun via groups.io:
>
> Dear all,
>
> Our two FTS system (Model TC-84) VT (picture attached) were broken. Could
> anybody share the manual for this system if you have it?
>
> Thanks in advance,
> --
> Weilin
>
> J-Star Research
>
> <NMR VT broken.jpg>
Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net Web: http://www.spin-doc.net
Representing:
Stelar Fast Field Cycling NMR Relaxometers (
http://www.stelar.it)
MestreLab NMR Software (
http://www.mestrelab.com)
Barthel HF Amplifiers (
http://www.barthel-hf.de)
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3743): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3743__;!!PvDODwlR4mBZyAb0!WsWHGEiuJUI-JhncH7Ao0E-LjNLYxp=
Pn0q9nqscMPoiH80J1hu5W1GLklPfBxruqt-O8snxTjAbu664-bKMx-A$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121059031=
/7559972__;!!PvDODwlR4mBZyAb0!WsWHGEiuJUI-JhncH7Ao0E-LjNLYxpPn0q9nqscMPoiH8=
0J1hu5W1GLklPfBxruqt-O8snxTjAbu667y5k_-8g$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Sep 02 2026 - 16:19:50 MST