From: Karel Klika via groups.io <karkli=utu.fi_at_groups.io>
Date: Fri, 9 May 2025 20:16:14 +0000

Dear all,

Many thanks to those that responded with suggestions – Bert, Thomas,
Hasan, Matthias, and Ivan – and also the personnel at Bruker (Lance
Ballard especially) for their assistance. It turns out that several weeks ago
we swapped over our antivirus program from Symantec to Defender and last week
was the first time we had rebooted the instrument since the change, and then
encountered problems (first strange error messages, then on a second attempt
to reboot, failure). Swapping back to Symantec – even though IT assured
us this was not the problem and they could not find any blocks in this respect
– and with multiple attempts at rebooting we were finally able to get
the system running again. Not a productive week I have to say. Anyway, I have
attached a list of things to check if rebooting is a problem with most suggestions
coming from Bruker.

Many thanks again to all.
K. Klika

________________________________________
> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Spin-Doc
> Sent: 06 May 2025 02:35:22
> To: main_at_ammrl.groups.io
> Cc: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] cannot reboot AVIII console

Hi,

Also be sure to check windows firewall settings as well as any external
ethernet connection.

Try a simple test: unplug the second Ethernet cable from the NMR PC, press
the Ipso reset button and see if it boots now.

I recently had a case where pinging 149.236.99.x got a reply from an actual
computer “out there” on the internet which interfered with
traffic to the console. The external IT company managing this customer’s
PCs had to come up with a routing script for their firewall/antivirus system to
prevent this address range from being directed to the outside world which prevented
the console from booting…

Cheers
Bert

-> Sent via mobile <-

Spin-Doc
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Web: http://www.spin-doc.net
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab, Stelar & Barthel HF



> Am 06.05.2025 um 00:58 schrieb Matthias Brandl via groups.io>:
>
> Dear Karel,
>
> If I remember correctly, the POST code C0 says that the (IPSO-internal) booting
> process went through, but the IPSO cannot find its operating system. The OS is
> normally provided by the PC via TFTP, so the correctly configured TFTP server
> needs to be running there.
>
> Unfortunately I don't have any experience with Windows spectrometer workstations,
> so I cannot really help any further...my first idea would be to check in the
> Task Manager whether some process with "tftp" in its name is running.
>
> Best,
> Matthias
>
>
>> On 05/05/2025 17:57, Karel Klika via groups.io wrote:
>> Dear all,
>> We are having problems getting our console to reboot, it is getting stuck
>> on "C0" at the postcode on the IPSO board when trying to reboot the console.
>> The top and bottom green LEDs are lit, the middle two are not. Otherwise all
>> other lights seems fine. We actually have a spare IPSO board and when that is
>> inserted the postcode only reads „00” and only the top LED is
>> lit. Attached are some of the error messages when TopSpin is opened, the first
>> half-dozen or so from the new IPSO board and the second half-dozen or so from
>> the old board. What to do to get the system working ?
>> We are running TS3.6.2 under Windows 10 and the system is an Avance III 400 single bay.
>> Any suggestions ? And thank you in advance.
>> Yours sincerely,
>> K. Klika
>
> --
> Dr. rer. nat. Matthias Brandl (he/him or they/them)
>
> Technical University of Munich
> TUM School of Natural Sciences - BNMRZ
> Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)
>
> Lichtenbergstr. 4 (postal)
> Ernst-Otto-Fischer-Str. 2 (physical)
> 85748 Garching b. Muenchen
>
> Phone +49 89 289-52617
>
> matthias.brandl_at_tum.de
> https://www.sattlerlab.de/
> https://www.bnmrz.org/
>
>

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2323): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2323__;!!PvDODwlR4mBZyAb0!RyA9cwKhCJzGTgrc_z9S32nXnbvAvi=
3DnG-GLO9zRSGwSMg0ez6eNOct1Xr40G9KlUywcqZNtN1xXCkN4z6567fR$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112632577=
/7559972__;!!PvDODwlR4mBZyAb0!RyA9cwKhCJzGTgrc_z9S32nXnbvAvi3DnG-GLO9zRSGwS=
Mg0ez6eNOct1Xr40G9KlUywcqZNtN1xXCkN4wM2hJOg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-





Received on Fri May 09 2025 - 18:16:32 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 01 2025 - 15:26:23 MST