From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Tue, 6 May 2025 07:35:22 +0800

Hi,

Also be sure to check windows firewall settings as well as any external
ethernet connection.

Try a simple test: unplug the second Ethernet cable from the NMR PC, press
the Ipso reset button and see if it boots now.

I recently had a case where pinging 149.236.99.x got a reply from an actual
computer “out there” on the internet which interfered with
traffic to the console. The external IT company managing this customer’s
PCs had to come up with a routing script for their firewall/antivirus system to
prevent this address range from being directed to the outside world which o
prevented the console from booting…

Cheers
Bert

-> Sent via mobile <-

Spin-Doc
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Web: http://www.spin-doc.net
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab, Stelar & Barthel HF



> Am 06.05.2025 um 00:58 schrieb Matthias Brandl via groups.io:
>
> Dear Karel,
>
> If I remember correctly, the POST code C0 says that the (IPSO-internal)
> booting process went through, but the IPSO cannot find its operating system.
> The OS is normally provided by the PC via TFTP, so the correctly configured
> TFTP server needs to be running there.
>
> Unfortunately I don't have any experience with Windows spectrometer workstations,
> so I cannot really help any further...my first idea would be to check in the
> Task Manager whether some process with "tftp" in its name is running.
>
> Best,
> Matthias
>
>
>> On 05/05/2025 17:57, Karel Klika via groups.io wrote:
>> Dear all,
>> We are having problems getting our console to reboot, it is getting stuck
>> on "C0" at the postcode on the IPSO board when trying to reboot the console.
>> The top and bottom green LEDs are lit, the middle two are not. Otherwise all
>> other lights seems fine. We actually have a spare IPSO board and when that
>> is inserted the postcode only reads „00” and only the top
>> LED is lit. Attached are some of the error messages when TopSpin is opened,
>> the first half-dozen or so from the new IPSO board and the second half-dozen
>> or so from the old board. What to do to get the system working ?
>> We are running TS3.6.2 under Windows 10 and the system is an Avance III 400 single bay.
>> Any suggestions ? And thank you in advance.
>> Yours sincerely,
>> K. Klika
>
> --
> Dr. rer. nat. Matthias Brandl (he/him or they/them)
>
> Technical University of Munich
> TUM School of Natural Sciences - BNMRZ
> Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)
>
> Lichtenbergstr. 4 (postal)
> Ernst-Otto-Fischer-Str. 2 (physical)
> 85748 Garching b. Muenchen
>
> Phone +49 89 289-52617
>
> matthias.brandl_at_tum.de
> https://www.sattlerlab.de/
> https://www.bnmrz.org/
>
>



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2298): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2298__;!!PvDODwlR4mBZyAb0!TnCdqQ_3XmH9feQ5UdMzaVlbkh_BPv=
coRq-VGEB-LeZ9K43tu3C8MVOCD_yeDsi9-U9oRQpccXjNwNYWaR_iZA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112632577=
/7559972__;!!PvDODwlR4mBZyAb0!TnCdqQ_3XmH9feQ5UdMzaVlbkh_BPvcoRq-VGEB-LeZ9K=
43tu3C8MVOCD_yeDsi9-U9oRQpccXjNwNa49Eu9dA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Mon May 05 2025 - 17:20:36 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 01 2025 - 15:26:23 MST