[AMMRL] AMMRL: Communications issues on AVII console

From: Richard Fitch via groups.io <richard.fitch=indstate.edu_at_groups.io>
Date: Tue, 10 Jun 2025 16:00:31 +0000

Dear Spinlanders,

We recently had our Avance II 400 lose communication with our workstation.
On power up, we get a few LINK and STATUS LEDs on the CCU flashing but then
nothing, leaving the first and second STATUS LEDs lit and the second LED on
the LINK lit. Resetting the CCU gives no changes except on the STATUS which
settles into the same first and second LEDs lit. The LEDs on the workstation
ethernet port are green and amber/yellow and static, so we have power and
connection but no communication.

Does anyone know the normal sequence of LEDs that should light up for both
the LINK and STATUS LEDs? Is there a reference for the indicators?

We have so far swapped the CCU and PSM2 as well as bypassed the router plugged
in directly to the workstation.

Any assistance would be greatly appreciated.

All the best,
Rick

Richard W. Fitch
Professor, Department of Chemistry and Physics
Councilor, Wabash Valley Section, American Chemical Society
Indiana State University
600 Chestnut Street
Science Building, Rm. S35E
Terre Haute, IN 47809
Phone: (812) 237-2244
Fax: (812) 237-2232

"Do an apparently simple experiment that gives you an important bit of information..."
Julius Axelrod

The love of money is the root of some evil. Fear and insecurity are the root
of most of it.

This email, and any attachments, thereto, is intended only for use by the
addressee(s) named herein and may contain privileged and/or confidential
information. If you are not the intended recipient of this email, you are
hereby notified that any dissemination, distribution or copying of this
email, and any attachments thereto, is strictly prohibited.



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2391): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2391__;!!PvDODwlR4mBZyAb0!UQO7D57ata_PZ_gZeiyDev5TJ41tKS=
dH_yYccG828pF-OFz2PYIBi7uWblJvltG3YWMhoA7HhbeOwpnogpPduLA8k_2R$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113572508=
/7559972__;!!PvDODwlR4mBZyAb0!UQO7D57ata_PZ_gZeiyDev5TJ41tKSdH_yYccG828pF-O=
Fz2PYIBi7uWblJvltG3YWMhoA7HhbeOwpnogpPduJ6La08q$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jun 10 2025 - 09:02:02 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST