Hi Rick

Here are my 2c worth. (I hope you have not been using any ethernet hub
between the workstation and the CCU/11 i.e. it is a direct connection i.e.
so-called 'eth1' connection). Other than that using a router or hub to
connect to the internet (so-called 'eth0' in old parlance) doesn't have
any bearing on the PC to CCU/11 connectivity.

   - You can use 'wireshark', which is a open source packet sniffer, to
   look at the TCP/IP communication packets between the two ends. This will
   help to see till what stage the diskless download and bootup proceeds.
   - If it was all working OK before, what changed now ? Was it a
   software upgrade i.e. OS or other ? Can you fall back to the older version
   and try ?
   - I can speak mostly for Linux systems and can you check if 'bootparamd'
   is running on your workstation when you try to boot up the CCU/11 ?
   - Not that the following will change spontaneously, but, can you make
   sure that the NFS v2 is active on your workstation ?

The wireshark dump will give a definitive foot print of how far you are
able to proceed.

Good Luck and Best Regards
Rajan

On Tue, Jun 10, 2025 at 9:02 AM Richard Fitch via groups.io wrote:

> Dear Spinlanders,
>
> We recently had our Avance II 400 lose communication with our
> workstation. On power up, we get a few LINK and STATUS LEDs on the CCU
> flashing but then nothing, leaving the first and second STATUS LEDs lit and
> the second LED on the LINK lit. Resetting the CCU gives no changes except
> on the STATUS which settles into the same first and second LEDs lit. The
> LEDs on the workstation ethernet port are green and amber/yellow and
> static, so we have power and connection but no communication.
>
> Does anyone know the normal sequence of LEDs that should light up for both
> the LINK and STATUS LEDs? Is there a reference for the indicators?
>
> We have so far swapped the CCU and PSM2 as well as bypassed the router
> plugged in directly to the workstation.
>
> Any assistance would be greatly appreciated.
>
> All the best,
>
> Rick
>
> Richard W. Fitch
> Professor, Department of Chemistry and Physics
>
> Councilor, Wabash Valley Section, American Chemical Society
> Indiana State University
> 600 Chestnut Street
> Science Building, Rm. S35E
> Terre Haute, IN 47809
> Phone: (812) 237-2244
> Fax: (812) 237-2232
>
> “Do an apparently simple experiment that gives you an important bit of
> information...”
> Julius Axelrod
>
> The love of money is the root of some evil. Fear and insecurity are the
> root of most of it.
>
> This email, and any attachments, thereto, is intended only for use by the
> addressee(s) named herein and may contain privileged and/or confidential
> information. If you are not the intended recipient of this email, you are
> hereby notified that any dissemination, distribution or copying of this
> email, and any attachments thereto, is strictly prohibited.
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2392): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2392__;!!PvDODwlR4mBZyAb0!Q3KQOzoWImMzywKllYl0Gn2f4765lq=
-1u5pRjRY1xf9UyvKfa4CStp8IZiNr6CLlVnJI7TNF4U9XxObz42j4UpMY$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113572508=
/7559972__;!!PvDODwlR4mBZyAb0!Q3KQOzoWImMzywKllYl0Gn2f4765lq-1u5pRjRY1xf9Uy=
vKfa4CStp8IZiNr6CLlVnJI7TNF4U9XxObz45IY2dhR$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jun 10 2025 - 11:33:05 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST