Re: [AMMRL] AMMRL: Communications issues on AVII console

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 11 Jun 2025 09:50:21 +0200

Hi,

one more thing to try: switch off the console for a while (to let it cool down),
switch it back on and see if you can now connect. If yes then it may be a
temperature-related issue like one or more fans stopped on top of the AQS
cardcage (worthwhile checking if yours already shows the running state of
AQS fans under ha - DRU-managed devices - card cage. Power supplies or other
boards may overheat as a consequence and result in undesired behavior…

Good luck hunting,
Bert

> Am 10.06.2025 um 18:00 schrieb Richard Fitch via groups.io:
>
> Dear Spinlanders,
>0
> We recently had our Avance II 400 lose communication with our workstation.
> On power up, we get a few LINK and STATUS LEDs on the CCU flashing but then
> nothing, leaving the first and second STATUS LEDs lit and the second LED on
> the LINK lit. Resetting the CCU gives no changes except on the STATUS which
> settles into the same first and second LEDs lit. The LEDs on the workstation
> ethernet port are green and amber/yellow and static, so we have power and
> connection but no communication.
>0
> Does anyone know the normal sequence of LEDs that should light up for both
> the LINK and STATUS LEDs? Is there a reference for the indicators?

> We have so far swapped the CCU and PSM2 as well as bypassed the router
> plugged in directly to the workstation.
>
> Any assistance would be greatly appreciated.
>
> All the best,
> Rick
>
> Richard W. Fitch
> Professor, Department of Chemistry and Physics
> Councilor, Wabash Valley Section, American Chemical Society
> Indiana State University
> 600 Chestnut Street
> Science Building, Rm. S35E
> Terre Haute, IN 47809
> Phone: (812) 237-2244
> Fax: (812) 237-2232
>
> “Do an apparently simple experiment that gives you an important bit of information...”
> Julius Axelrod
>
> The love of money is the root of some evil. Fear and insecurity are the root of most of it.
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software (http://www.mestrelab.com
Barthel HF Amplifiers (http://www.barthel-hf.de


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2401): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2401__;!!PvDODwlR4mBZyAb0!QcVXsx41z0fZW5pkOswk80Z_LLgzQB=
Rsh8Uu3KPiydzRunCOdKTq2hppktAB2QtAhjchI8UdSfO9AF9J2OGmmA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113572508=
/7559972__;!!PvDODwlR4mBZyAb0!QcVXsx41z0fZW5pkOswk80Z_LLgzQBRsh8Uu3KPiydzRu=
nCOdKTq2hppktAB2QtAhjchI8UdSfO9AF_-Ie3ziw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Jun 11 2025 - 03:26:59 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST