Re: [AMMRL] Varian VT Connector Fix

From: Jessi Baughman via groups.io <jessi_baughman=goodyear.com_at_groups.io>
Date: Tue, 10 Jun 2025 12:07:23 -0700

Hi Bruce,

In the files section ( https://ammrl.groups.io/g/main/files )
of ammrl.group.io, there are a number of hardware manuals. The
pinout you are looking for is on page 10 of the Highland VT
controller manual. Note that two of the pins need to be type T
alloys, not the standard copper of replacement DB 9 connectors.

https://ammrl.groups.io/g/main/files/Hardware*20Manuals/VarianAgilent*20Hardware*20Manuals/L900_Highland_VT_Controller_2009.pdf*page=10

>From the manual:

>
> *4.2.1 Probe Connector J101 (Rev C Units)*
> The rear-panel 9-pin probe connector J101 is a female, 9-pin D-SUB type.
> The two thermocouple-input pins
> are of type 'T' alloys, and the mating connector must have corresponding
> thermocouple alloy crimp pins.
> Pinout is as follows:
> * Thermocouple + Copper wire/pin alloy
> * Ground
> * Thermocouple -Constantin wire/pin alloy
> * n/c
> * n/c
> * Heater highside drive
> * Heater lowside drive (ground return)
> * Safety R TD low side
> * Safety R TD high side
>

*Jessi Baughman*
Scientist Staff, Analytical Science * Akron
The Goodyear Tire & Rubber Company
142 Goodyear Blvd, Akron, OH 44305


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2394): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2394__;!!PvDODwlR4mBZyAb0!XBgrvPgouwzerqBrqJI50YqkHIglv2=
zvClmJ6OQbVB9hgWWf1gzEeSIsPGVooYVDiPILeWJM_wXri6LK1m1DAfJwJurG$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112057683=
/7559972__;!!PvDODwlR4mBZyAb0!XBgrvPgouwzerqBrqJI50YqkHIglv2zvClmJ6OQbVB9hg=
WWf1gzEeSIsPGVooYVDiPILeWJM_wXri6LK1m1DAVztKxEc$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jun 10 2025 - 12:07:32 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST