Hi Everyone
I feel like this is just another thread having a problem with IPSO!
So after a power outage I can not talk to IPSO. This is an AVIII running
topspin 3.7 on Alma 9.
We have made sure tftp, and dhcpd are running properly. I have reinstalled
diskless. numerous power cycles of console and computer.
IPSO code is 2C or 2E depending on if I tried to reset IPSO. I can ping other
components so Ethernet and switch are good.
Any other suggestions?
Thanks
Greg
Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry #3255
580 Ross St.
College Station, TX 77843-3255
gpwylie_at_tamu.edu
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3176): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3176__;!!PvDODwlR4mBZyAb0!SWtayvmMRSn_t28eP9a7dg75ImrfmN=
iZigIWqdKNEJKUdTMuB3Y9d7M6a-wDwayMq-57bVpHPPG73KWq0e5ZjRGiyg9N$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117642432=
/7559972__;!!PvDODwlR4mBZyAb0!SWtayvmMRSn_t28eP9a7dg75ImrfmNiZigIWqdKNEJKUd=
TMuB3Y9d7M6a-wDwayMq-57bVpHPPG73KWq0e5ZjRDJoIXL$
Mute #bruker:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!SWtayvmMRSn_t28eP9a7dg75ImrfmNiZigIWqd=
KNEJKUdTMuB3Y9d7M6a-wDwayMq-57bVpHPPG73KWq0e5ZjV0djIeR$
Mute #hardware:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!SWtayvmMRSn_t28eP9a7dg75ImrfmNiZig=
IWqdKNEJKUdTMuB3Y9d7M6a-wDwayMq-57bVpHPPG73KWq0e5ZjTDh-7GG$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Feb 04 2026 - 12:40:44 MST