Re: [AMMRL] AVIII running Topspin 3.7 on Alma IPSO issue #Bruker #Hardware

From: Alex Greenwood via groups.io <greenwa2=ucmail.uc.edu_at_groups.io>
Date: Thu, 5 Feb 2026 03:37:15 +0000

Hi Greg,

For my AVIII HD/TS3.8 in Alma Linux 9.6 system, something like this happens
every time I need to power down. The only thing that works for me is to
reset the IPSO with a pencil once the console is booted, then install Topspin
again in parallel with the install I’m using. During the installation
it asks to reboot the IPSO (I say yes!)—this has been a reliable fix so far.

Alex


Alex Greenwood (he/him)
Manager, NMR Facility
University of Cincinnati
409 Old Chemistry
Office: 513-556-9211; Cell: 412-952-8044
https://www.artsci.uc.edu/natural-sciences/chemistry/core-facilities/nuclear-magnetic-resonance-facility.html

> From: main_at_ammrl.groups.io On Behalf Of gregory.wylie via groups.io
> Sent: Wednesday, February 4, 2026 1:59 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] AVIII running Topspin 3.7 on Alma IPSO issue #Bruker #Hardware


Hi Everyone

I feel like this is just another thread having a problem with IPSO!

So after a power outage I can not talk to IPSO. This is an AVIII running
topspin 3.7 on Alma 9.

We have made sure tftp, and dhcpd are running properly. I have reinstalled
diskless. numerous power cycles of console and computer.

IPSO code is 2C or 2E depending on if I tried to reset IPSO. I can ping other
components so Ethernet and switch are good.

Any other suggestions?

Thanks
Greg

Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry #3255
580 Ross St.
College Station, TX 77843-3255
gpwylie_at_tamu.edu
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3178): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3178__;!!PvDODwlR4mBZyAb0!S4NH-mkDnrUhkpujMS3aarX1KQBxra=
K_gc2kAVnjXXCzbe7hchgu_ryjCEnGAHNntVpUKoS99bhXkuKFRtKhcIiCTW8z$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117642432=
/7559972__;!!PvDODwlR4mBZyAb0!S4NH-mkDnrUhkpujMS3aarX1KQBxraK_gc2kAVnjXXCzb=
e7hchgu_ryjCEnGAHNntVpUKoS99bhXkuKFRtKhcGMxI83_$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!S4NH-mkDnrUhkpujMS3aarX1KQBxraK_gc2kAV=
njXXCzbe7hchgu_ryjCEnGAHNntVpUKoS99bhXkuKFRtKhcMGXnQ2G$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!S4NH-mkDnrUhkpujMS3aarX1KQBxraK_gc=
2kAVnjXXCzbe7hchgu_ryjCEnGAHNntVpUKoS99bhXkuKFRtKhcD4BAZAC$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Feb 05 2026 - 05:11:00 MST

This archive was generated by hypermail 2.4.0 : Sun Mar 01 2026 - 13:47:31 MST