Hi Greg,
I just pulled up my PhoenixBIOS 4.0 Release 6.0. If they are correct:
2Ch " Ram failure on address line ****"
2Eh "Ram failure on date bits **** of low byte of memory bus"
Good luck on fixing the problem.
Best,
Simon
________________________________
> From: main_at_ammrl.groups.io on behalf of gregory.wylie via groups.io
> Sent: Wednesday, February 4, 2026 12:58 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] AVIII running Topspin 3.7 on Alma IPSO issue #Bruker #Hardware
Hi Everyone
I feel like this is just another thread having a problem with IPSO!
So after a power outage I can not talk to IPSO. This is an AVIII running
topspin 3.7 on Alma 9.
We have made sure tftp, and dhcpd are running properly. I have reinstalled =
diskless. numerous power cycles of console and computer.
IPSO code is 2C or 2E depending on if I tried to reset IPSO. I can ping other
components so Ethernet and switch are good.
Any other suggestions?
Thanks
Greg
Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry #3255
580 Ross St.
College Station, TX 77843-3255
gpwylie_at_tamu.edu
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3182): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3182__;!!PvDODwlR4mBZyAb0!SYu7nyFa12UUlvG1Zh-wbfx94plw6X=
D803usF1dWFn_IYgP3ydhVa2Qe9UZJ-TXPWX8btNECgsA_domxetswq9_V$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117642432=
/7559972__;!!PvDODwlR4mBZyAb0!SYu7nyFa12UUlvG1Zh-wbfx94plw6XD803usF1dWFn_IY=
gP3ydhVa2Qe9UZJ-TXPWX8btNECgsA_domxev1t_95z$
Mute #bruker:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!SYu7nyFa12UUlvG1Zh-wbfx94plw6XD803usF1=
dWFn_IYgP3ydhVa2Qe9UZJ-TXPWX8btNECgsA_domxemZD3H5z$
Mute #hardware:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!SYu7nyFa12UUlvG1Zh-wbfx94plw6XD803=
usF1dWFn_IYgP3ydhVa2Qe9UZJ-TXPWX8btNECgsA_domxesjTvC_B$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Feb 05 2026 - 09:41:20 MST