Re: [AMMRL] AVIII running Topspin 3.7 on Alma IPSO issue #Bruker #Hardware

From: ik54 via groups.io <ik54=cornell.edu_at_groups.io>
Date: Thu, 05 Feb 2026 10:00:27 -0800

Hi Greg,

This happens regularly on our AVIII as well with a Windows PC. The solution
that has worked for us is to turn off everything in the console and restart
just the IPSO before powering on anything else.

To shut down: Run 'ha' from Topspin and perform soft shutdown of IPSO and BSMS.
Power off BSMS, IPSO, AQX rack, amplifiers and sample changer, then the external
power switch.

To restart: reboot computer, turn on external power switch, and power on IPSO. Don't
power on anything else until IPSO gets to 94. If it doesn't get there on the first
try, keep trying by pushing the reset button on the IPSO.

Restarting the Bruker dhcp and tftp server services may help. When IPSO makes it to
94, power on everything else. I usually go top to bottom, but it probably doesn't
make a difference. Wait a few minutes then start Topspin and 'cf'.

It's possible that turning off the BSMS might be enough. It generates a LOT of
network traffic, which may prevent the IPSO from establishing communications with
the host.

Good luck,
Ivan

--
Ivan Keresztes, Ph.D.
Director, NMR Facility
Department of Chemistry and Chemical Biology, Cornell University Baker Laboratory, Ithaca, NY 14853.
phone: (607) 255 0709, fax: (607) 255 4137, email: ik54_at_cornell.edu
(he/him/his)

nmr.chem.cornell.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3183): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3183__;!!PvDODwlR4mBZyAb0!WlBJgJJHCn9BAXM3NYCa9wX4GOFSo8=
pHWEioELH4BKYsTQ407sH4HjcTnnKZMnunG_vI3t_NM_A2huXoY1lJIZFbu7mD$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117642432=
/7559972__;!!PvDODwlR4mBZyAb0!WlBJgJJHCn9BAXM3NYCa9wX4GOFSo8pHWEioELH4BKYsT=
Q407sH4HjcTnnKZMnunG_vI3t_NM_A2huXoY1lJIVTgseW6$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!WlBJgJJHCn9BAXM3NYCa9wX4GOFSo8pHWEioEL=
H4BKYsTQ407sH4HjcTnnKZMnunG_vI3t_NM_A2huXoY1lJIZwkbRa9$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!WlBJgJJHCn9BAXM3NYCa9wX4GOFSo8pHWE=
ioELH4BKYsTQ407sH4HjcTnnKZMnunG_vI3t_NM_A2huXoY1lJIdlnRpwo$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Feb 05 2026 - 10:24:27 MST

This archive was generated by hypermail 2.4.0 : Sun Mar 01 2026 - 13:47:31 MST