Re: [AMMRL] PTS Adaptor Card for VNMRS console #agilent #Hardware #Troubleshooting #varian #VarianAgilent

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 26 Feb 2025 21:33:54 +0100

Hi,

please check the tiny "F1" fuse in the bottom right corner - they have a
tendency of breaking (= open connection). Simply bridge it with a thin
piece of wire and it may work again.

Are you sure that all individual modules inside the VFS failed? Might be the
internal reference generator unit as well - if you want you can send it to
me and I can at least check it or even repair it.

Cheers
Bert

> From: main_at_ammrl.groups.io on behalf of brcherry via groups.io
> Sent: Wednesday, February 26, 2025 7:19:10 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] PTS Adaptor Card for VNMRS console #agilent #Hardware #Troubleshooting #varian #VarianAgilent
>
>0
> Hi,
> I'm still trying to keep our Varian VNMRS 800 MHz system up and running.
> We've had issues with our frequency synthesizers that have knocked out 2
> of our 4 channels. All four channels in a VFS unit have failed. I have a
> pile of PTS units that still function. However, my two spare interface cards
> seem to have failed. (0191411603). These are the cards that convert the
> VNMRS "ethernet" to the PTS multi pin interface port. Without them i can't
> leverage the PTS frequency Synthesizer to regain 3 or 4 channel function.
>
> Does anyone have any that they are willing to part with?
>
> Thank you,
> Brian Cherry
> Arizona State University
> Magnetic Resonance Research Center
>
>
>
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software (http://www.mestrelab.com
Barthel HF Amplifiers (http://www.barthel-hf.de



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2069): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2069__;!!PvDODwlR4mBZyAb0!S3ZAMUyCjalsGfKVYrodEzJ37pHIdP=
hGemva1lKU6dAcD05WIDvAgh6OD69THlTy813mGnBiLjnfgI2texxqPg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111402897=
/7559972__;!!PvDODwlR4mBZyAb0!S3ZAMUyCjalsGfKVYrodEzJ37pHIdPhGemva1lKU6dAcD=
05WIDvAgh6OD69THlTy813mGnBiLjnfgI28yG8Fuw$
Mute #agilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!S3ZAMUyCjalsGfKVYrodEzJ37pHIdPhGemva=
1lKU6dAcD05WIDvAgh6OD69THlTy813mGnBiLjnfgI3bI3JNqA$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!S3ZAMUyCjalsGfKVYrodEzJ37pHIdPhGem=
va1lKU6dAcD05WIDvAgh6OD69THlTy813mGnBiLjnfgI2glaoIwQ$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!S3ZAMUyCjalsGfKVYrod=
EzJ37pHIdPhGemva1lKU6dAcD05WIDvAgh6OD69THlTy813mGnBiLjnfgI2InGVJTg$
Mute #varian:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/varian__;!!PvDODwlR4mBZyAb0!S3ZAMUyCjalsGfKVYrodEzJ37pHIdPhGemva1l=
KU6dAcD05WIDvAgh6OD69THlTy813mGnBiLjnfgI072hm-tA$
Mute #varianagilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/m=
ain/mutehashtag/varianagilent__;!!PvDODwlR4mBZyAb0!S3ZAMUyCjalsGfKVYrodEzJ3=
7pHIdPhGemva1lKU6dAcD05WIDvAgh6OD69THlTy813mGnBiLjnfgI2iW9VGTQ$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!S3ZAMUyCjalsGfKVY=
rodEzJ37pHIdPhGemva1lKU6dAcD05WIDvAgh6OD69THlTy813mGnBiLjnfgI3OfMtAzA$ [wa=
lt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Feb 26 2025 - 12:38:55 MST

This archive was generated by hypermail 2.4.0 : Fri Feb 28 2025 - 15:51:13 MST