Re: [AMMRL] PTS Adaptor Card for VNMRS console #agilent #Hardware #Troubleshooting #varian #VarianAgilent

From: David Rovnyak via groups.io <drovnyak=bucknell.edu_at_groups.io>
Date: Wed, 26 Feb 2025 15:25:30 -0500

had to deal with multiple VFS failures in past too; agree this is likely
one of the key things that can keep vnmrs consoles going. if a broader
solution emerges it could help a lot of folks. -David

> ------------------------------
> *From:* main_at_ammrl.groups.io on behalf of brcherry> via groups.io
> *Sent:* Wednesday, February 26, 2025 7:19:10 PM
> *To:* main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> *Subject:* [AMMRL] PTS Adaptor Card for VNMRS console #agilent #Hardware #Troubleshooting #varian #VarianAgilent
>
>
> Hi,
> I'm still trying to keep our Varian VNMRS 800 MHz system up and running.
> We've had issues with our frequency synthesizers that have knocked out 2 of
> our 4 channels. All four channels in a VFS unit have failed. I have a
> pile of PTS units that still function. However, my two spare interface
> cards seem to have failed. (0191411603). These are the cards that convert
> the VNMRS "ethernet" to the PTS multi pin interface port. Without them i
> can't leverage the PTS frequency Synthesizer to regain 3 or 4 channel
> function.
>
> Does anyone have any that they are willing to part with?
>
> Thank you,
> Brian Cherry
> Arizona State University
> Magnetic Resonance Research Center
>
>
>
>
>
>
--
---------------------------------------------------------------------------
David Rovnyak, Professor of Chemistry, Bucknell Univ., Lewisburg, PA 17837
BlueSky : NMR_Guy <https://bsky.app/profile/nmrguy.bsky.social>
Research: http://www.facstaff.bucknell.edu/drovnyak/
Bucknell Chemistry: http://www.departments.bucknell.edu/chemistry/
Pascal: "I would like to have written a shorter letter, but I didn't have the time"
---------------------------------------------------------------------------


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2068): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2068__;!!PvDODwlR4mBZyAb0!RTogSIJWlgv0Eqa2Nwr3EXR-WEm5Fe=
cJPUAzK8cWLzd-hpULP8I57LVzpeOTVWu293j0SVigXLNqlKbAKD7XeTkquPnr$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111402897=
/7559972__;!!PvDODwlR4mBZyAb0!RTogSIJWlgv0Eqa2Nwr3EXR-WEm5FecJPUAzK8cWLzd-h=
pULP8I57LVzpeOTVWu293j0SVigXLNqlKbAKD7XeczkLeYr$
Mute #agilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!RTogSIJWlgv0Eqa2Nwr3EXR-WEm5FecJPUAz=
K8cWLzd-hpULP8I57LVzpeOTVWu293j0SVigXLNqlKbAKD7Xea9bA-Hp$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!RTogSIJWlgv0Eqa2Nwr3EXR-WEm5FecJPU=
AzK8cWLzd-hpULP8I57LVzpeOTVWu293j0SVigXLNqlKbAKD7Xeas4Oo_V$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!RTogSIJWlgv0Eqa2Nwr3=
EXR-WEm5FecJPUAzK8cWLzd-hpULP8I57LVzpeOTVWu293j0SVigXLNqlKbAKD7XecQy-HUj$
Mute #varian:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/varian__;!!PvDODwlR4mBZyAb0!RTogSIJWlgv0Eqa2Nwr3EXR-WEm5FecJPUAzK8=
cWLzd-hpULP8I57LVzpeOTVWu293j0SVigXLNqlKbAKD7XeYEp1uAG$
Mute #varianagilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/m=
ain/mutehashtag/varianagilent__;!!PvDODwlR4mBZyAb0!RTogSIJWlgv0Eqa2Nwr3EXR-=
WEm5FecJPUAzK8cWLzd-hpULP8I57LVzpeOTVWu293j0SVigXLNqlKbAKD7XeVrT9ZLS$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Feb 26 2025 - 12:25:48 MST

This archive was generated by hypermail 2.4.0 : Fri Feb 28 2025 - 15:51:13 MST