Re: [AMMRL] Bruker communication / Syntax error from AQS3 (AV2+)

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 26 Feb 2025 22:01:17 +0100

Dear Mike,

thanks for the bit of clarification - in that case it would be a bit odd if
the RXAD can be talked to by Unitool *through* the DRU (it is required to
be connected and I get an error when talking to aqs/SBSB16 while the DRU is
unplugged from ethernet) but the DRU cannot be talked to by Unitool…
Actually the AV3 hardware manual says "The boards within the AQS that are
directly accessible via Unitool are: SGU, REF, RXAD, internal amplifiers."
So I think we shouldn't get the syntax errors from the REF1 and AMP (but
maybe DRU

In the end our main problem is the failure of "cf" during HPPR detection -
which may or may not have to do with these errors…

In the meantime I tried out a couple of more things:
- swapped to the second ethernet card in the (Windows) computer
- swapped the PSD3 card
- checked all power supplies in the backside for any defective or almost defective
capacitors (measured ESD of all of them - all look OK)
- all power supply LEDs are green
- tried to start the console with super basic setup (unplugged both SGUs,
unplugged IPSO from ethernet and still got the Unitool syntax error from REF1 but
connection OK to RXAD) - just to see whether some board "eats" too much power
- reduced HPPR stack to one single preamp
- reinstalled Topspin (3.7.0)

So far to no avail - cf is still stuck at the detection of the last HPPR/2
module (now the only module = P1)

Cheers

Bert


> Am 26.02.2025 um 20:25 schrieb Michael Groves via groups.io:
>
> Hi Bert,
>
> I'm pretty sure the DRU doesn't talk to UniTool--you have to use HA to go
> to the service webpage. And I'm pretty sure the amplifier is kind of a
> simple dumb device and you can't talk to it either. And I'm not sure
> communication is possible with the REF board either. It's been a while
> since I messed with that vintage spectrometer but I think there are some
> bits you just can't talk to. Having said that, it is entirely possible that
> I'm completely wrong, but it feels like I'm right. I'm pretty sure there's
> no firmware or anything on the REF board and you just plug it in and it
> works or not. And I know there is firmware on the full size amplifiers but
> I don't think the BLABB has anything like that.
>
> If this is a system with an IPSO then all that stuff is taken care of through
> the DRU service webpage. If it's a system with a CCU in the AQS rack instead
> then the DRU still has a service webpage and some of those other bits will
> talk to UniTool.
>
> But enough rambling. It sounds to me like things might be working okay.
>
> Cheers,
> Mike
>
> On Wednesday, February 26, 2025 at 11:21:51 AM MST, Spin-Doc wrote:
>
>
> Hi,
>
> at a customer place I observe some strange behaviour:
>
> In Unitool I can talk to some boards (SGU1, SGU2, RXAD) but not to some others (DRU, REF1, AMP1):
>
> aqs:
> RXAD (SBSB 16): OK
> SGU1 (SBSB 36): OK
> SGU2 (SBSB 37): OK
> AMP1 (SBSB 54): ERROR B020 from BLA1 Slot 9 Rack1: Syntax error
> REF1 (SBSB 52): ERROR B020 from REF1 Slot 3 Rack1: Syntax error
> DRU (SBSB 44): Communication error detected
>
> hppr:
> all OK
>
> When trying to connect to DRU the "transfer wheel" isn't "rotating" at all
> whereas it does for RXAD/SGUs. No menu displayed.
>
> Also "cf" keeps hanging infinitely at le last line of HPPR detection:
> read BIS from HPPR/2 module P4 …
>
> I have replaced the ethernet switch and cables (didn't seem to be the issue
> because a flood ping of 5000 packages to 149.236.99.89 went through with
> flying colours).
> Replaced the DRU board with another one - no effect.
> Replaced the HPPR cover unit (cf hanging was found before the Unitool comms errors).
>
> Any ideas? Main power supply maybe?
>
> Best regards
> Bert
>
>
> Magnetic resonance equipment, service, training & more
> Selling & buying used NMR equipment (all vendors)
> Dr. rer. nat. Bert Heise
> Owner, Applications Scientist
>
> Spin-Doc Bert Heise
> Kiefernweg 13
> 58239 Schwerte
> Germany
> Mobile: +49 171 5135315
> Email: bert.heise_at_spin-doc.net
> Web: http://www.spin-doc.net
>
> Representing:
> Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
> MestreLab NMR Software (http://www.mestrelab.com)
> Barthel HF Amplifiers (http://www.barthel-hf.de)
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2071): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2071__;!!PvDODwlR4mBZyAb0!XHD-h4HESLZXcXgoIX7_AkvCTD-888=
SIf_68wgwrukyctd5VFu-Jw4zv2VPsqeuuGIiAbjupfHMT4Osebz3hZQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111401280=
/7559972__;!!PvDODwlR4mBZyAb0!XHD-h4HESLZXcXgoIX7_AkvCTD-888SIf_68wgwrukyct=
d5VFu-Jw4zv2VPsqeuuGIiAbjupfHMT4Ou2XEU6Fg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Feb 26 2025 - 13:31:30 MST

This archive was generated by hypermail 2.4.0 : Fri Feb 28 2025 - 15:51:13 MST