Re: [AMMRL] Bruker SampleCase error

From: Kenneth Sharp-Knott <kknott_at_vt.edu>
Date: Tue, 16 Apr 2024 21:22:13 +0000

Hi Perry,

I'm not sure if it's the exact same error, but I can tell you that I experienced
something very similar after upgrading to TS3.7.0. Fortunately, i have an older
AVII that also has a SampleCase (that cannot be upgraded) so I was able to do a
side-by-side comparison.

After the update to 3.7.0, and importantly the irreversible firmware update,
even the slightest bump of the red bar under the transport tube will shutdown the
SampleCase. Previously, you had to really press into it and hold it for a second
before this would happen. After all, it's meant to detect an overly long NMR tube
from breaking, not to prevent a finger from scuffing a nail. I don't recall the
error on screen, but the symptom was the same in that this shut down the SampleCase
and the automation.

After much whining accompanied by frequent complaints to anyone who would listen, I
was ultimately told that this was a requirement by Pharma (safety compliance) and
would not be rolled back, ie. deal with it. Grumble, grumble, grumble....

Our solution was a big sign saying DON'T TOUCH THE RED BAR and student users have since
been more careful about removing or putting in the samples in that area.

Apparently, it is impossible to add this as an optional toggle for Pharma/Academia. This
is an example of why users are often reticent to update their TS versions.

I think this is likely the same issue you are dealing with.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1147): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1147__;!!PvDODwlR4mBZyAb0!RKDL_1FJFvIxfykdRkvQjb15W1IvqL=
jy1TdEh391kjdvdODlCGBVSSFuuPu8bXFxoblmzZra9z9MPQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105565488=
/7559972__;!!PvDODwlR4mBZyAb0!RKDL_1FJFvIxfykdRkvQjb15W1IvqLjy1TdEh391kjdvd=
ODlCGBVSSFuuPu8bXFxoblmzZr8vjWgJg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Apr 16 2024 - 14:22:24 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:17 MST