Re: [AMMRL] Bruker SampleCase error

From: Thusitha Jayasundera via groups.io <thusitha.jayasundera=bc.edu_at_groups.io>
Date: Tue, 16 Apr 2024 17:13:37 -0400

Hi Perry--I had this exact issue on 9/15/2023, and reached out to the group
at the time. The only difference was that we upgraded from Topspin 4.1.1 to
4.3 at the time. The solution was, unfortunately, to revert back to 4.1.1,
and there have been zero issues since. There were other suggestions from
the group though (which you should be able to find via a search), but if
all else fails, reverting back to the old version of Topspin should work.

Good luck!

TJ.


Thusitha Jayasundera, Ph.D.
Director, Magnetic Resonance Center
Boston College
Merkert 105
2609 Beacon Street
Chestnut Hill MA 02467-3860

Tel: (617) 552-6984 (office)
Fax: (617) 552-2705 <617-552-2705>

Website:
https://www.bc.edu/content/bc-web/research/sites/vice-provost-for-research/about/research-facilities/nmr-center


On Tue, Apr 16, 2024 at 5:05 PM Perry Pellechia via groups.io wrote:

> Hello All,
> I am working with Bruker on a fairly intermittent error we are getting on
> a AV-IIIHD system running with a SampleCase (ambient, 24 samples). The
> error is showing up on average every 4-7 days (100-200 samples). This
> started shortly after upgrading the host workstation to Topspin 3.7.0 (new
> HP Z4G5 workstation with Win 11). We were not seeing this before the
> upgrade (TS 3.6.1 on Win 10). I am not necessarily seeking a solution (it
> would be fantastic if there is one w/o a software patch) but looking for
> anyone else who might be seeing the same or similar error.
>
> The message we are seeing (similar on both Topspin and IconNMR) is:
>
> “WebService Error: 1000 STOP_CHANGER SampleTransporter: System not ready
> to accept new command”
>
> With the error, the sample changer goes “red" and automation stops. This
> certainly seems like a software bug and the issue has been upgraded to
> second tier support. Any additional information could be helpful for
> coming up with a software/firmware fix.
>
> Since this is our primary walkup instrument you probably can see that this
> is fairly inconvenient. Other than this, the 3.7.0 upgrade has gone very
> well on our 4 systems. This was probably the most trouble free software
> upgrade I have applied and I was happy to retire 10-12 year old computers.
> However, this system stopping automation remains the only significant issue.
>
> Thanks for any similar reports or suggestions.
>
> Perry.
> --
>
> Perry J. Pellechia
> Director of NMR Services
> Research Professor
> University of South Carolina
> Department of Chemistry and Biochemistry
>
>
>
>
>
> --
> Perry.
> --
>
> Perry J. Pellechia
> Director of NMR Services
> Research Professor
> University of South Carolina
> Department of Chemistry and Biochemistry
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1146): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1146__;!!PvDODwlR4mBZyAb0!V6A8ekncan094WnnU25dwRH19oRWL0=
W_Q9OKtO_Tu455qqnPyH01SK7eo83U9i_nO5fvD3313GvNhnuBQsBQgsol86EH$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105565488=
/7559972__;!!PvDODwlR4mBZyAb0!V6A8ekncan094WnnU25dwRH19oRWL0W_Q9OKtO_Tu455q=
qnPyH01SK7eo83U9i_nO5fvD3313GvNhnuBQsBQgiqSCNgj$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Apr 16 2024 - 14:13:57 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:17 MST