[AMMRL] AlmaLinux 9 workstation from scratch

From: Anne Carroll via groups.io <carrol50=miamioh.edu_at_groups.io>
Date: Tue, 29 Sep 2026 14:05:58 -0700

Hello,

I would like to know if any of you have successfully built a spectrometer
workstation from scratch using the instructions Bruker provides?

AlmaLinux 9 TopSpin Installation
( https://repo-bbio.bruker.com/AlmaLinux/9/SoftwareInstallationRequirements.html )
Icon has been sporadically freezing on our 400 MHz instrument (AVANCE Neo,
TS 4.5.0, Icon 6.2.2). I think it started after we upgraded both the TS and
Icon software in April 2025. The computer runs CentOS 7, which reached end-of-life
in June 2024, and arrived with the spectrometer in 2020 preconfigured by Bruker.
I haven't figured out any consistent trigger for the freeze.

TopSpin continues to work most of the time, but I usually have to restart
TS to get Icon operating again.

Apparently, TopSpin 4.5 is not validated for CentOS 7. Based on the savelogs
I shared, the only suggestion the Bruker team in Europe made was that I update
my operating system from CentOS 7 to Alma 9.

I now have a computer running AlmaLinux 9 with enough PCI ports to connect
to both the spectrometer and the internet. I believe I have followed the
installation and configuration instructions correctly, but the instructions
don't give any way to check before connecting the computer to a console and
installing TopSpin. This instrument is a workhorse in my department, and I'm
scared I will break it by corrupting the Codemeter license or something.

I will appreciate any words of encouragement or warning,

Anne Carroll
--
Anne M. Carroll, Ph.D. (she/her)
NMR/MS Instrumentation Specialist
Department of Chemistry and Biochemistry
Miami University
Oxford, Ohio USA
513-529-3163 Office
678-637-6739 Cell
carrol50_at_miamioh.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3807): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3807__;!!PvDODwlR4mBZyAb0!UEJNK7IXNFxLv08AMBHyxKf5zKwTaW=
l4FUuDfUP7D3Em3w2GQn-P4ceZZ246G78jAQldb3uHlQRvdJ6LBX7W70lakEOt$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121500420=
/7559972__;!!PvDODwlR4mBZyAb0!UEJNK7IXNFxLv08AMBHyxKf5zKwTaWl4FUuDfUP7D3Em3=
w2GQn-P4ceZZ246G78jAQldb3uHlQRvdJ6LBX7W7zlYPjC_$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Sep 29 2026 - 16:37:01 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST