Re: [AMMRL] AlmaLinux 9 workstation from scratch

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Wed, 30 Sep 2026 13:44:33 +0000 (UTC)

Hi Anne,

One thing you might try would be to install any CentOS updates you haven't
yet.  I believe you have to tell the updater to look at vault.centos.org
instead of mirror.centos.org (been a while since I had to mess with that but
I think it's correct).  Sometimes the newest versions of topspin run better
when the OS is up-to-date (at least as much as possible as well).

If you want to see if it connects to the spectrometer, the best way to test
it is just to do it.  It should be simple enough to switch back if it
doesn't.  The NEO spectrometers aren't quite as fiddly to get communicating
as the older versions.  If you've got the spectrometer connection set to
149.236.99.1/24 (alma) or 149.236.99.1 with subnet mask 255.255.255.0 (centos)
it should be able to talk to things.  If not, especially if you're running
TS4.5 on both, switching back shouldn't be a problem.  You'll probably
get a "server mismatch detected" message and it should overwrite the servers
on the console (and likely change nothing) and then go back to normal operation. 
I know of at least a couple of cases where the user had to try a few times to get
things communicating properly and was able to put the original computer back
without any problems so that their users could use the instrument.  

Regarding the license, I think the most you can do is cause yourself some delays
and inconvenience.  If you REALLY REALLY REALLY foul things up then you
might have to request a new license but I'm not sure I've ever seen that happen. 
The most likely things that we see are that you either have to ask us to release
the license or that you have to delete the current CmContainer and generate a new one. 
Neither of these is too complicated.  If you simply release the license
from the existing computer and move it over to the new one that process shouldn't
cause any problems and should only take a couple of minutes each time.  

TL;DR--switching between computers shouldn't be too complicated.  And it's
VERY unlikely that you'll mess up the license to the point that it's unrecoverable. 
I think you should just block out some time to swap computers and see what happens. 
And you can always go back to the original one.

Cheers,Mike

On Tuesday, September 29, 2026 at 05:36:58 PM MDT, Anne Carroll via groups.io wrote:

Hello,

I would like to know if any of you have successfully built a spectrometer
workstation from scratch using the instructions Bruker provides? 
AlmaLinux 9 TopSpin InstallationIcon has been sporadically freezing on our 400 MHz
instrument (AVANCE Neo, TS 4.5.0, Icon 6.2.2). I think it started after we
upgraded both the TS and Icon software in April 2025. The computer runs
CentOS 7, which reached end-of-life in June 2024, and arrived with the
spectrometer in 2020 preconfigured by Bruker. I haven't figured out any
consistent trigger for the freeze. TopSpin continues to work most of the
time, but I usually have to restart TS to get Icon operating again.
Apparently, TopSpin 4.5 is not validated for CentOS 7. Based on the savelogs
I shared, the only suggestion the Bruker team in Europe made was that I update
my operating system from CentOS 7 to Alma 9.I now have a computer running Alma
Linux 9 with enough PCI ports to connect to both the spectrometer and the internet.
I believe I have followed the installation and configuration instructions
correctly, but the instructions don't give any way to check before connecting
the computer to a console and installing TopSpin. This instrument is a workhorse
in my department, and I'm scared I will break it by corrupting the Codemeter
license or something. I will appreciate any words of encouragement or warning,

Anne Carroll

--
Anne M. Carroll, Ph.D. (she/her)
NMR/MS Instrumentation Specialist
Department of Chemistry and Biochemistry
Miami University
Oxford, Ohio USA
513-529-3163 Office
678-637-6739 Cell
carrol50_at_miamioh.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3810): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3810__;!!PvDODwlR4mBZyAb0!T4A59Nsam4FuipNRSAHxYFoxdWwSYr=
roDh5iNXqlk0eIJorWznaS55HwxbTROhvTOka95vO1mQZh2B75-8Gbvw6YXd9r$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121500420=
/7559972__;!!PvDODwlR4mBZyAb0!T4A59Nsam4FuipNRSAHxYFoxdWwSYrroDh5iNXqlk0eIJ=
orWznaS55HwxbTROhvTOka95vO1mQZh2B75-8Gbv_O-TPIW$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Sep 30 2026 - 07:13:00 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST