HI Anne
We have rebuilt 8 NMR computers to use Alma 9, 2 of them Varian systems the
rest Bruker. I can not say we followed the Bruker instructions to the letter
but they were helpful. It has been several years since this was done and we
have only have one Bruker that I can say "locks up". This is a Neo 500MHz
installed in 2023 and it will routinely freeze or not do things properly.
We end up having to reboot or power off the console to "fix" the issues.
As for IconNMR and sample changers, I can not say anything that happens with
our sample changers or Icon is Alma related. Things can "lock up" but usually
because the sample changer had an issue. We also don't see any issues that
we did not see when using CentOS.
We do have a mix of consoles and Topspin versions:
Avance III (2010) Topspin 3.7.0
Avance III Nano Bay (2013) Topspin 3.7.0
2 x NEO (2018) Topspin 4.3.0
Neo Solids (2020) Topspin 4.3.0
Neo (2023) Topspin 4.5.0
Neo-X (install in 2 weeks so we will see what happens)
Greg
Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry #3255
580 Ross St.
College Station, TX 77843-3255
gpwylie_at_tamu.edu
979.458.0705 ( tel:(979)%20458-0705 ) (voice)
979.845.4719 ( tel:(979)%20845-4719 ) (fax)
http://nmr.tamu.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3811): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3811__;!!PvDODwlR4mBZyAb0!U1YJTc5tFMKqO6Ps9dNRc94HRZTmOv=
9DiwqCMAyBP5qWXBydac2JM64DjpLLbANLjOOJZx1FjcilbngSoCE8gWcWlnUV$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121500420=
/7559972__;!!PvDODwlR4mBZyAb0!U1YJTc5tFMKqO6Ps9dNRc94HRZTmOv9DiwqCMAyBP5qWX=
Bydac2JM64DjpLLbANLjOOJZx1FjcilbngSoCE8gXqqNwaS$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Sep 30 2026 - 07:13:17 MST