Re: [AMMRL] AlmaLinux 9 workstation from scratch

From: Bob Berno via groups.io <bberno=mcmaster.ca_at_groups.io>
Date: Wed, 30 Sep 2026 09:14:35 -0400

Hello Anne,

I have not seen IconNMR "freezing" as you describe.  Instead, I have
seen that IconNMR will suddenly report that no probe is installed. Or
that it has lost communication with the BSMS.  And this will happen
between consecutive experiments running one after the other! Note,
TopSpin still recognizes the probe and it can still communicate with the
BSMS.  It is just IconNMR that stops because of these reported issues.
When this happens we find that all we need to do is close TopSpin and
start it again.  We might run "edprobe" to make sure it "detects" the
probe, but even that isn't always necessary.  Our workaround is to exit
TopSpin and restart at least once a week, and often we can get through a
week without this happening.

We have seen this behaviour only with TopSpin 4.5.0 systems with
SampleCase or SampleCasePlus.  We have a system running TopSpin 4.1.4
with a SampleJet that can run IconNMR non-stop for months without error.

All of this is to say that I think your issues are may be due to TopSpin
4.5.0 and not your OS install.

You might want to try to reinstall only the IconNMR portion of TopSpin.

If there is one small library file related to the IconNMR operation got
corrupted, then reinstalling might fix that.

Alternatively, you could just practice the workaround of closing and
restarting IconNMR once or twice a week.

When you say that IconNMR freezes, I assume you mean the desktop
interface is frozen, correct?  Have you enabled the remote web
interface?  I'm wondering if the IconWeb interface remains active while
the desktop is frozen.  If that is the case, maybe it is an issue with

the graphics driver, or maybe Java?

Good luck,
Bob.

On 2026-09-29 5:05 p.m., Anne Carroll via groups.io wrote:
>
> Hello,
> I would like to know if any of you have successfully built a
> spectrometer workstation from scratch using the instructions Bruker
> provides?
> AlmaLinux 9 TopSpin Installation
> https://repo-bbio.bruker.com/AlmaLinux/9/SoftwareInstallationRequirements.html
> Icon has been sporadically freezing on our 400 MHz instrument (AVANCE
> Neo, TS 4.5.0, Icon 6.2.2). I think it started after we upgraded both
> the TS and Icon software in April 2025. The computer runs CentOS 7,
> which reached end-of-life in June 2024, and arrived with the
> spectrometer in 2020 preconfigured by Bruker. I haven't figured out
> any consistent trigger for the freeze. TopSpin continues to work most
> of the time, but I usually have to restart TS to get Icon operating again.
> Apparently, TopSpin 4.5 is not validated for CentOS 7. Based on the
> savelogs I shared, the only suggestion the Bruker team in Europe made
> was that I update my operating system from CentOS 7 to Alma 9.
> I now have a computer running AlmaLinux 9 with enough PCI ports to
> connect to both the spectrometer and the internet. I believe I have
> followed the installation and configuration instructions correctly,
> but the instructions don't give any way to check before connecting the
> computer to a console and installing TopSpin. This instrument is a
> workhorse in my department, and I'm scared I will break it by
> corrupting the Codemeter license or something.
> I will appreciate any words of encouragement or warning,
> Anne Carroll
> --
> Anne M. Carroll, Ph.D. (she/her) NMR/MS Instrumentation Specialist
> Department of Chemistry and Biochemistry Miami University Oxford, Ohio
> USA513-529-3163 Office 678-637-6739 Cell carrol50_at_miamioh.edu
>

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga hand
Haudenosaunee Nations, and within the lands protected by the "Dish With One
Spoon" wampum agreement.



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3808): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3808__;!!PvDODwlR4mBZyAb0!R53-3rrqMugfxivzxSZtAghn2XLhBu=
MWuU72uS7tEcRGrgSjb4pXeo1t7PO2-zN2SYIIL8gK12YNzVVhu7Y3JtITa9po$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121500420=
/7559972__;!!PvDODwlR4mBZyAb0!R53-3rrqMugfxivzxSZtAghn2XLhBuMWuU72uS7tEcRGr=
gSjb4pXeo1t7PO2-zN2SYIIL8gK12YNzVVhu7Y3JgjRRvz0$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Sep 30 2026 - 06:18:07 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST