This sounds like the heater for the reference frequency crystal was off for
a while making the frequency source unstable until it has time to warm up
again. With my Inova 500 from that time period, I always wait about 6 hr.
before trying to lock and shim to give it sufficient time to stabilize.
Bruce D. Ray
________________________________
> From: main_at_ammrl.groups.io on behalf of Monika Ivancic via groups.io
> Sent: Monday, September 14, 2026 4:56 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [External] [AMMRL] lock'n'shim out of control on a Varian Unity/Inova 500
Dear fellow spinlanders,
I am trying to keep this Varian Unity/Inova 500 from 1997 alive, and have had
various issues along the way. The spectrometer doesn't get much use, and at
some point will need to be retired. As of late, I've been presented with a
new issue.
The day I came back from vacation, I believe things on the spect looked normal,
however just before lunch I had noticed that the "Xmas tree" lights were out.
I then checked the console lights and they were out as well, although the power
supply (bottom R side) was still running. So, I powered down the entire spectrometer,
and waited 10-15 minutes, then brought it back up. It came up OK, with all the
lights behaving normally. However, when it came to locking on a solvent, the
lock was behaving erratically. It was tricky finding the on resonance position
on a CDCl3 sample, because the lock signal would jump from negative to positive.
Perhaps this was due to lock phase being off, especially since when the console
would be brought down and then powered back up, typically the lock phase would change.
I then used a sucrose in h2o/d2o sample, found the initial on-resonance position at
z0=5179, and once locked, it regulated the lock fairly decently. However, I
then proceeded with a 3D gradient autoshimming, which then moved the z0 to 5380-ish,
which is quite a bit off from the original value (I often would find a different z0
after 3D autoshimming, but it would only be 10 or so units off). And the lock
behaved a bit more erratically at this point.
Then I tried the lineshape sample (1%CHCl3 in acetone-d6), the lock behaved somewhat
normal at a higher lock power, but when I brought the lock power down to 14, the
appropriate value, prior to locking the lock level would bounce from as high as ~83
to as low as ~35. Once locking on the signal, with z0=1843, the lock bounced a
little, and I proceeded to do 1D gradient autoshimming (having to recable first,
since we're detecting 2H in this case). Unfortunately this autoshimming didn't
improve the lock level and resulted in the lock level bouncing quite a bit when
'regulated'. I decided to repeat this 1D autoshimming, but after this try, I had
to relock, and now z0=2083, and the lock level is holding in place when 'regulated'.
However, when I went to acquire the spectrum, I found 'lock wiggles' in every peak,
and the peaks were broad.
Finally, I decided to 'reset' the console, in hopes this would improve the situation,
and starting with the sucrose in h2o/d2o sample. Initially the lock was behaving
more or less normally, however when I found on resonance it was at z0=3903 (wildly
different from a few days before with the same sample). The regulated lock wasn't
bad, only fluctuating between 73 and 69. But then I went and did 3D gradient
Autoshimming, but then the x-shim was taken out of range, AND I had to relock and
the new z0=4733. The x1 and y1 shims had been taken way out, so they needed a
large manual correction. However once acquiring the spectrum, the lock was wildly
oscillating, between 82% and 69%.
Not sure what else I could try, perhaps bringing the console back down and then
back up, but at some point I'm just wasting my time, so I thought I'd share this
issue with you all to see what you think? Likely I need some new boards for the
locking and shimming mechanism. And yes, a new console would be wonderful! ;-)
With gratitude,
Monika
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3787): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3787__;!!PvDODwlR4mBZyAb0!VrtYpvlTNgO_y3W5XQfUf8h4-gvrOK=
O49GM4rXHmL5afgjkkmjtDNwr_QYw6HwzKrWPympJUnQeOPxdOR__pmg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121251037=
/7559972__;!!PvDODwlR4mBZyAb0!VrtYpvlTNgO_y3W5XQfUf8h4-gvrOKO49GM4rXHmL5afg=
jkkmjtDNwr_QYw6HwzKrWPympJUnQeOPxdtNU13Sg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Sep 14 2026 - 18:50:49 MST