All—This issue began with a PTS that failed because of dried-out
capacitors. Replacing the caps restored the PTS, but now when running any scan
of any nucleus, it finishes the first transient scan but hangs with "FID:1 CT: 1"
in the status box. After about 30 seconds it times out and returns to idle.
wft on the command line says "no data in file". However, subsequent scans
show an error message stating that the "experiment is in progress. Use
"aa" to abort it and then ga" even though it was idle. I've run testControllers,
and all 7 tests passed. I've reinstalled vnmrj4.2A with the same status result.
Any help would be appreciated. -- Robert Harker
-=-=-=-=-=-=-=-=-=-=-=-