From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Mon, 14 Sep 2026 23:27:38 +0200

Dear Monika,

that all sounds quite odd indeed… Could you possibly check the following:

- get a multimeter and measure all voltages at the bottom right corner of your
console - all voltages should be 0.1-0.2 V higher than the printed value
(5.1-5.2 V instead of 5.0, let alone 4.9 V)

- send via email/post here a screenshot of the Lock Scan" of a 100% D2O sample
(the 90:10% H2O:D2O sucrose and CDCl3 samples aren't great when chasing lock
issues because of their rather low S/N in 2H). Does it show a spike or glitch
at the beginning of the on-resonant lock FID? That might indicate an issue
with the Lock XCVR board (defective mixer chip)

- am I right in assuming your console doesn't have the lock relay in order to
observe deuterium without recabling? Is it a 10.5 or 20 MHz Intermediate Frequency
Inova? Maybe send me some photos of the connector side of the "magnet leg"
(via private email, see my signature below).

- please check whether the fans in the shim power supply, main power supply
and RF amplifier are all working - maybe something is overheating.

- Is a remote login via Anydesk or Teamviewer possible by any chance?

Best regards
Bert

> Am 14.09.2026 um 22:56 schrieb Monika Ivancic via groups.io:
>
> Dear fellow spinlanders,
>
> I am trying to keep this Varian Unity/Inova 500 from 1997 alive, and have
> had various issues along the way. The spectrometer doesn't get much use,
> and at some point will need to be retired. As of late, I've been presented
> with a new issue.
>
> The day I came back from vacation, I believe things on the spect looked normal,
> however just before lunch I had noticed that the "Xmas tree" lights were out.
> I then checked the console lights and they were out as well, although the power
> supply (bottom R side) was still running. So, I powered down the entire
> spectrometer, and waited 10-15 minutes, then brought it back up. It came up
> OK, with all the lights behaving normally. However, when it came to locking
> on a solvent, the lock was behaving erratically. It was tricky finding the
> on resonance position on a CDCl3 sample, because the lock signal would jump
> from negative to positive. Perhaps this was due to lock phase being off,
> especially since when the console would be brought down and then powered back
> up, typically the lock phase would change.
>
> I then used a sucrose in h2o/d2o sample, found the initial on-resonance
> position at z0=5179, and once locked, it regulated the lock fairly decently.
> However, I then proceeded with a 3D gradient autoshimming, which then moved
> the z0 to 5380-ish, which is quite a bit off from the original value (I often
> would find a different z0 after 3D autoshimming, but it would only be 10 or
> so units off). And the lock behaved a bit more erratically at this point.
>
> Then I tried the lineshape sample (1%CHCl3 in acetone-d6), the lock behaved
> somewhat normal at a higher lock power, but when I brought the lock power
> down to 14, the appropriate value, prior to locking the lock level would
> bounce from as high as ~83 to as low as ~35. Once locking on the signal,
> with z0=1843, the lock bounced a little, and I proceeded to do 1D gradient
> autoshimming (having to recable first, since we're detecting 2H in this
> case). Unfortunately this autoshimming didn't improve the lock level and
> resulted in the lock level bouncing quite a bit when 'regulated'. I decided
> to repeat this 1D autoshimming, but after this try, I had to relock, and
> now z0=2083, and the lock level is holding in place when 'regulated'.
> However, when I went to acquire the spectrum, I found 'lock wiggles' in
> every peak, and the peaks were broad.
>
> Finally, I decided to 'reset' the console, in hopes this would improve the
> situation, and starting with the sucrose in h2o/d2o sample. Initially the
> lock was behaving more or less normally, however when I found on resonance
> it was at z0=3903 (wildly different from a few days before with the same
> sample). The regulated lock wasn't bad, only fluctuating between 73 and
> 69. But then I went and did 3D gradient Autoshimming, but then the x-shim
> was taken out of range, AND I had to relock and the new z0=4733. The x1
> and y1 shims had been taken way out, so they needed a large manual correction.
> However once acquiring the spectrum, the lock was wildly oscillating,
> between 82% and 69%.
>
> Not sure what else I could try, perhaps bringing the console back down and
> then back up, but at some point I'm just wasting my time, so I thought I'd
> share this issue with you all to see what you think? Likely I need some
> new boards for the locking and shimming mechanism. And yes, a new console
> would be wonderful! ;-)
>
> With gratitude,
> Monika
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3786): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3786__;!!PvDODwlR4mBZyAb0!RExTu_9VsQgIPfkYLZFuQgu14Yu2GQ=
OrN5VdoOwlbBTcXcEsFwlQrk7jfEGzO6FhkyNGoB11sTiN7GBoZS-QkA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121251037=
/7559972__;!!PvDODwlR4mBZyAb0!RExTu_9VsQgIPfkYLZFuQgu14Yu2GQOrN5VdoOwlbBTcX=
cEsFwlQrk7jfEGzO6FhkyNGoB11sTiN7GBcVzY11A$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Mon Sep 14 2026 - 14:29:33 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST